7.2 To-Dos before switching off
Before switching off the device, the following points must be observed:
- Correct stowage of the probe
- Checking the fine dust filters
- Emptying the water trap
WARNING
The hot probe can cause damage to the instrument or to the transport
system.
WARNING
Highly contaminated filters can affect the volumetric flow in the gas path and
therefore the measurement
WARNING
For instruments without automatic condensate emptying, condensation
water from the condensation tank can flow into the gas path inside the
instrument and damage the sensors.
(Only for devices WITHOUT automatic condensate emptying.)
1
ecom-EN3
Let the probe cool down before stowing it!
Check the condition of the particle filter.
Replace the particle filter when the contamination of the filter
is equivalent to a soot number of approx. 2 - 3.
Empty the condensate trap before transporting the
instrument.
Emptying the water trap as follows:
1. Pull the drain hose
2. Drain the condensate
3. Re-attach the drain hose
2
3