Summary of Contents for Hanil SeroSpin II

  • Page 1 Operating Manual SeroSpin II Clinical Centrifuge Date of Purchase Serial No. Place of purchase...
  • Page 2 SeroSpin II Medical Device Copyright ⓒ 2018 Hanil Scientific Inc. All rights reserved. Contact Us If you have and questions, contact Hanil Scientific Inc. or place of purchase. +82-2-3472-0727 Inquiry: info@ihanil.com Order: sales@ihanil.com Tech. support: techsupport@ihanil.com The contents in this operating manual are subject to change for device improvement.
  • Page 3: Table Of Contents

    CONTENTS General Considerations------------------------------------------------- Safety Transport & Storage Safety label Electric safety information Product description------------------------------------------------------- Structure -5 Delivery package -5 2.3 Technical specifications -6 Installation ---------------------------------------------------------------- Packing Inspection Installation -7 3.2.1 Selecting the location -7 3.2.2 Power connetion -8 Loading and removal of the rotor - 8 Operation ---------------------------------------------------------------- Operation panel...
  • Page 4: General Considerations

    2. Check the voltage to be used, before connecting the centrifuge to the power source. 3. Only use rotors, parts, and accessories provided by Hanil Scientific Inc. Hanil Scientific Inc. is not responsible for damages of the device and accidents caused by using parts and accessories not recommended.
  • Page 5: Transport & Storage -4

    1. General Considerations 1.2 Transport & Storage • Only lift and transport the • The device and the accessories device with sufficient number of may only be stored in dry rooms helpers. -Storage- -Transport- Ambient temperature 5℃~40℃ Ambient temperature -10℃~40℃ Maximum relative humidity 30%~85% Maximum relative humidity 10%~90% Air pressure 500~1060hpa...
  • Page 6: Product Description

    2.1 Structure ① ⑤ ② ③ ④ ① Door ② Chamber ③ Display&Control panel ④ Power switch ⑤ RPM measuring window 2.2 Delivery package ① SeroSpin II Main body(incl. a rotor) ② Power Cable ③ Operating manual 5 l www.ihanil.com...
  • Page 7: Technical Specifications -6

    2. Product description 2.3 Technical Specifications Max. RPM 3,400 rpm Max. RCF 1,118 xg Max. Capacity 12 x 75 mm tube ACC/DEC ramps 5/6 steps Time control < 100 min, pulse, continuous Program memory Noise level ≤ 60 dB Imbalance cutoff Dimension (W x D x H, mm) 293 x 380 x 242 Weight without rotor...
  • Page 8: Installation

    3. Installation 3.1 Packing Inspection •Check packing conditions carefully, before unpacking. •Contact Hanil Scientific Inc. immediately if damages found. •Check the delivery for completeness. •You can get contact details on packing boxes and the back of the manual. 3.2 Installation 3.2.1 Selecting the location...
  • Page 9: Installation -7

    2. Switch the centrifuge on using the power switch on the right side of the device. 3.3 Loading and removal of the rotor •Only use rotors, parts, and accessories provided by Hanil Scientific Inc. •Do not use scatched or cracked rotor in use.
  • Page 10: Operation

    4. Operation 4.1 Operation Panel Button Function Set the speed or runtime Move the cursor lerf or right Time Accel Set the speed or runtime Decel Save the value Call/Save Save or call up a program Door Start centrifugation Start/Stop/ Stop centrifugation Short www.ihanil.com...
  • Page 11: Loading Tubes -10

    4. Operation 4.2 Opening and closing the Door •The door can only be opened if the centrifuge is switched on. •Do not reach with your fingers between the housing and door. •Close the door completely before operation. •Opening the door Press the Door ( ) button.
  • Page 12: Setting Rpm/Rcf -11

    4. Operation 4.4 Setting RPM/RCF ① Press the Enter( ) button. ②Press Left ( ) or Right ( ) button to move the cursor to RPM or RCF mode. ▶ RPM mode: RPM display flashes. ▶ RCF mode: RCF display flashes. ③Press Up ( ) or Down ( ) button until the required RPM/RCF displayed.
  • Page 13: Setting Accelation/Deceleration Rate -12

    4. Operation 4.6 Setting Accelation/Deceleration rate ① Press the Enter( ) button. ②Press Left ( ) or Right ( ) button to move the cursor to Accel/Decel. ▶ Accelation setting mode:Accel display flashes. ▶ Deceleration setting mode:Decel display flashes. ③Press Up ( ) or Down ( ) button until the required number is displayed.
  • Page 14: Start The Centrifugation Run -13

    4. Operation 4.8 Start the centrifugation run ① Press Start/Stop ( ) button. 4.9 End the centrifugation run ① Press Start / Stop ( ) button to stop centrifugation. 4.10 Short spin The [Short] button is for when press & hold, to accelerate up to maximum speed and decelerate when the button is not pressed.
  • Page 15: Maintenance

    ②Rotate the shaft with your hand to make sure it turns smoothly ③Use the stopwatch to check that the time setting is correct ④If you find any damages, do not use the device. Contact Hanil Scientific Inc. 5.2 Cleaning • Before cleaning the centrifuge, be sure to switch off the device and disconnect the power cord.
  • Page 16: Troubleshooting

    6. Troubleshooting 6.1 General errors Problem Recommended Action Check the power cord connection. Power failure Check the power fuse of the device. Device cannot be started Check the lid is closed completely. Lid cannot be opened Press the ‘Lid open button’. Remove the dirt at the door latch and close the lid.
  • Page 17: Error Messages -16

    6. Troubleshooting 6.2 Error messages If you cannot remedy an error with recommended actions, please contact Hanil Scientific Inc. Error Problem Cause Recommended Action code • Asymmetrical placement of tubes. • Check the tightness of rotor • Irregular volume of sample in loading.
  • Page 18 6. Troubleshooting 6.2 Error messages Error Problem Cause Recommended Action code Low voltage •Motor applied voltage decreases • Switch power off. Error abnormally during operation. Sensing • Rpm sensor failure • Restart the device. error • Occurs when communica communication between Main and •...
  • Page 19: Emergency Lid Open -18

    6. Troubleshooting 6.3 Emergency lid open If the centrifug door cannot be opened, you can activate the emergency open manually. 1. Wait for rotor to stop before activating the emergency open. 2. Insert a wrench into the emergency release hole and turn it clockwise until the lid is released.
  • Page 20: Rotor

    7. Rotor 7.1 Fixed angle rotors 19 l www.ihanil.com...
  • Page 21 Memo...
  • Page 22 Memo...
  • Page 23 Memo...
  • Page 24 Hanil Scientific Inc. 16 Arayukro, Gimpo 10136, Rep. of KOREA T. 02-3452-8965 info@ihanil.com www.ihanil.com...