  • Page 1 Instructions for use INTRA K-LUX 200 - 1.000.6717 | INTRA K-LUX 200 F - 1.002.1912...
  • Page 2 Distributed by: Manufacturer: KaVo Dental GmbH Kaltenbach & Voigt GmbH Bismarckring 39 Bismarckring 39 D-88400 Biberach D-88400 Biberach Phone +49 7351 56-0 www.kavo.com Fax +49 7351 56-1488...
  • Page 3: Table Of Contents

    Instructions for use INTRA K-LUX 200 - 1.000.6717 | INTRA K-LUX 200 F - 1.002.1912 Contents Contents 1 User instructions ........................ 5 2 Safety..........................7 2.1 Infection hazard ......................7 2.2 Explosion hazard ......................7 2.3 Electrical shock......................7 2.4 Technical condition......................7 2.5 Electromagnetic fields ....................
  • Page 4 Instructions for use INTRA K-LUX 200 - 1.000.6717 | INTRA K-LUX 200 F - 1.002.1912 Contents 9 Warranty terms and conditions..................23 4 / 26...
  • Page 5: User Instructions

    © Copyright by KaVo Dental GmbH KaVo Original Factory Repair In the event of a repair, please ship your product to the KaVo Original Factory Repair using www.ka- vobox.com. KaVo Technical Service...
  • Page 6 Instructions for use INTRA K-LUX 200 - 1.000.6717 | INTRA K-LUX 200 F - 1.002.1912 1 User instructions Please note the electronic instructions for use Note: observe accompanying documents EAC conformity mark (Eurasian Conformity) GOST R certification Transportation and storage conditions...
  • Page 7: Safety

    Instructions for use INTRA K-LUX 200 - 1.000.6717 | INTRA K-LUX 200 F - 1.002.1912 2 Safety | 2.1 Infection hazard 2 Safety The instructions for use are a component of the product and must be read care- fully prior to use and be accessible at all times.
  • Page 8: Electromagnetic Fields

    Instructions for use INTRA K-LUX 200 - 1.000.6717 | INTRA K-LUX 200 F - 1.002.1912 2 Safety | 2.5 Electromagnetic fields ▶ If the following defects occur, stop working and have the service personnel carry out repair work: ▪ Malfunctions ▪...
  • Page 9: Service And Repair

    Instructions for use INTRA K-LUX 200 - 1.000.6717 | INTRA K-LUX 200 F - 1.002.1912 2 Safety | 2.8 Service and repair 2.8 Service and repair Repairs and servicing may only be performed by trained service personnel. The following persons are authorised to do this: ▪...
  • Page 10: Product Description

    Instructions for use INTRA K-LUX 200 - 1.000.6717 | INTRA K-LUX 200 F - 1.002.1912 3 Product description | 3.1 Purpose – Intended use 3 Product description INTRA K-LUX 200 (Mat. no. 1.000.6717) INTRA K-LUX 200 F (Mat. no. 1.002.1912) 3.1 Purpose –...
  • Page 11 Instructions for use INTRA K-LUX 200 - 1.000.6717 | INTRA K-LUX 200 F - 1.002.1912 3 Product description | 3.2 Technical data Motor electronics Material nos. 0.641.1020, 0.641.1120, 0.641.1140, 0.696.0161 and 1.004.4317 Motor voltage rating 26 V DC Motor speed rating...
  • Page 12: Operation With High-Pressure Lamp

    Instructions for use INTRA K-LUX 200 - 1.000.6717 | INTRA K-LUX 200 F - 1.002.1912 3 Product description | 3.3 Transportation and storage conditions 3.2.1 Operation with high-pressure lamp Operating voltage of the high-pres- max. 3.2 V DC sure lamp High-pressure lamp power max.
  • Page 13: Start Up And Shut Down

    Instructions for use INTRA K-LUX 200 - 1.000.6717 | INTRA K-LUX 200 F - 1.002.1912 4 Start up and shut down 4 Start up and shut down WARNING Hazard from non-sterile products. Infection hazard for dentist and patient. ▶ Prior to initial startup and after each use, reprocess the product and accessories.
  • Page 14: Operation

    Instructions for use INTRA K-LUX 200 - 1.000.6717 | INTRA K-LUX 200 F - 1.002.1912 5 Operation | 5.1 Connecting the motor to the supply hose 5 Operation Note At the beginning of each workday, the water-conducting systems should be...
  • Page 15: Removing Attachment Instruments From The Motor

    Instructions for use INTRA K-LUX 200 - 1.000.6717 | INTRA K-LUX 200 F - 1.002.1912 5 Operation | 5.4 Removing attachment instruments from the motor NOTICE Attaching and removing instruments during rotation. Property damage to the product. ▶ Do not attach or remove the instruments during rotation.
  • Page 16: Checking For Malfunctions And Troubleshooting

    Instructions for use INTRA K-LUX 200 - 1.000.6717 | INTRA K-LUX 200 F - 1.002.1912 6 Checking for malfunctions and troubleshooting | 6.1 Checking for malfunctions at initial start-up 6 Checking for malfunctions and troubleshooting 6.1 Checking for malfunctions at initial start-up CAUTION Danger from hot lamp.
  • Page 17: Replacing The O-Rings

    Instructions for use INTRA K-LUX 200 - 1.000.6717 | INTRA K-LUX 200 F - 1.002.1912 6 Checking for malfunctions and troubleshooting | 6.2 Troubleshooting ▶ Insert a new lamp in the recess so that the contact surfaces mate with those in the holder.
  • Page 18: Reprocessing Steps In Accordance With Iso 17664

    Instructions for use INTRA K-LUX 200 - 1.000.6717 | INTRA K-LUX 200 F - 1.002.1912 7 Reprocessing steps in accordance with ISO 17664 | 7.1 Preparation at the site of use 7 Reprocessing steps in accordance with ISO 17664 7.1 Preparation at the site of use WARNING Hazard from contaminated products.
  • Page 19: Manual Internal Cleaning

    Instructions for use INTRA K-LUX 200 - 1.000.6717 | INTRA K-LUX 200 F - 1.002.1912 7 Reprocessing steps in accordance with ISO 17664 | 7.2 Manual Reprocessing ▶ Wipe the motor body and handpiece coupling with a dry cloth and remove any residual oil.
  • Page 20: Manual Internal Disinfection

    Instructions for use INTRA K-LUX 200 - 1.000.6717 | INTRA K-LUX 200 F - 1.002.1912 7 Reprocessing steps in accordance with ISO 17664 | 7.3 Automated reprocessing ▶ Spray the disinfectant on a cloth, then wipe down the medical device and allow the disinfectant to act according to the instructions of the disinfectant manufacturer.
  • Page 21: Packaging

    Instructions for use INTRA K-LUX 200 - 1.000.6717 | INTRA K-LUX 200 F - 1.002.1912 7 Reprocessing steps in accordance with ISO 17664 | 7.5 Packaging 7.5 Packaging Not applicable. Note Only the removable motor sleeve can be sterilised in an autoclave at up to 7.6 Sterilisation...
  • Page 22: Tools And Consumables

    Instructions for use INTRA K-LUX 200 - 1.000.6717 | INTRA K-LUX 200 F - 1.002.1912 8 Tools and consumables 8 Tools and consumables Available from dental suppliers. Material summary Mat. no. High-pressure bulb 1.002.2928 O-ring 0.200.6120 Airflow measuring tube 0.411.4441...
  • Page 23 Instructions for use INTRA K-LUX 200 - 1.000.6717 | INTRA K-LUX 200 F - 1.002.1912 9 Warranty terms and conditions 9 Warranty terms and conditions The following warranty conditions apply to this KaVo medical device: KaVo provides the end customer with a warranty of proper function and guar-...