KaVo SMARTmatic S10 Instructions For Use Manual

KaVo SMARTmatic S10 Instructions For Use Manual

Hide thumbs Also See for SMARTmatic S10:

Advertisement

Quick Links

Instructions for use
SMARTmatic S10 REF 1.011.6730
SMARTmatic S10 K REF 1.011.6731
SMARTmatic S10 S REF 1.011.6732

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SMARTmatic S10 and is the answer not in the manual?

Questions and answers

Summary of Contents for KaVo SMARTmatic S10

  • Page 1 Instructions for use SMARTmatic S10 REF 1.011.6730 SMARTmatic S10 K REF 1.011.6731 SMARTmatic S10 S REF 1.011.6732...
  • Page 2 Distributed by: Manufacturer: KaVo Dental Corporation Kaltenbach & Voigt GmbH 11729 Fruehauf Drive Bismarckring 39 Charlotte, NC 28273 USA D-88400 Biberach Phone: 847 550 6800 www.kavo.com Fax: 847 550 6825...
  • Page 3: Table Of Contents

    Instructions for use SMARTmatic S10 REF 1.011.6730 SMARTmatic S10 K REF 1.011.6731 SMARTmatic S10 S REF 1.011.6732 Table of contents Table of contents 1 User instructions............................. 5 2 Safety ................................7 2.1 Infection hazard ............................. 2.2 Technical condition ..........................2.3 Accessories and combination with other equipment................
  • Page 4 Instructions for use SMARTmatic S10 REF 1.011.6730 SMARTmatic S10 K REF 1.011.6731 SMARTmatic S10 S REF 1.011.6732 Table of contents 8 Tools and consumables..........................26 9 Terms and conditions of warranty ........................27 4 / 30...
  • Page 5: User Instructions

    Instructions for use SMARTmatic S10 REF 1.011.6730 SMARTmatic S10 K REF 1.011.6731 SMARTmatic S10 S REF 1.011.6732 1 User instructions 1 User instructions Dear User, congratulations on purchasing this KaVo quality product. By following the instructions below you will be able to work smoothly, economically and safely.
  • Page 6 Instructions for use SMARTmatic S10 REF 1.011.6730 SMARTmatic S10 K REF 1.011.6731 SMARTmatic S10 S REF 1.011.6732 1 User instructions CE mark according to Medical Devices Directive EC 93/42 Consult instructions for use Caution: Consult instructions for use EAC conformity mark (Eurasian Conformity)
  • Page 7: Safety

    Instructions for use SMARTmatic S10 REF 1.011.6730 SMARTmatic S10 K REF 1.011.6731 SMARTmatic S10 S REF 1.011.6732 2 Safety | 2.1 Infection hazard 2 Safety The instructions for use are a component of the product and must be read carefully prior to use and be accessible at all times.
  • Page 8: Accessories And Combination With Other Equipment

    Instructions for use SMARTmatic S10 REF 1.011.6730 SMARTmatic S10 K REF 1.011.6731 SMARTmatic S10 S REF 1.011.6732 2 Safety | 2.3 Accessories and combination with other equipment ▶ Do not use motors for unauthorized purposes. 2.3 Accessories and combination with other equipment Use of un-authorized accessories on the device or un-authorized modifications to the device can lead to injury.
  • Page 9: Description Of The Product

    Instructions for use SMARTmatic S10 REF 1.011.6730 SMARTmatic S10 K REF 1.011.6731 SMARTmatic S10 S REF 1.011.6732 3 Description of the product | 2.5 Service and repair 3 Description of the product SMARTmatic S10 (Mat. No. 1.011.6730) SMARTmatic S10 K (Mat. No. 1.011.6731) SMARTmatic S10 S (Mat.
  • Page 10: Intended Use

    Instructions for use SMARTmatic S10 REF 1.011.6730 SMARTmatic S10 K REF 1.011.6731 SMARTmatic S10 S REF 1.011.6732 3 Description of the product | 3.1 Intended use 3.1 Intended use Indications for Use The SMARTmatic handpieces are intended for the removal of carious material, reduc-...
  • Page 11: Transportation And Storage Conditions

    Instructions for use SMARTmatic S10 REF 1.011.6730 SMARTmatic S10 K REF 1.011.6731 SMARTmatic S10 S REF 1.011.6732 3 Description of the product | 3.3 Transportation and storage conditions Amount of air spray min. 1.5 Nl/min (at 2 bar) To use the spray: ▶...
  • Page 12: Startup And Shut-Down

    Instructions for use SMARTmatic S10 REF 1.011.6730 SMARTmatic S10 K REF 1.011.6731 SMARTmatic S10 S REF 1.011.6732 4 Startup and shut-down | 4.1 Checking the amount of water for the S10 S 4 Startup and shut-down WARNING Hazard from nonsterile products Infection danger to the care provider and patient.
  • Page 13: Operation

    Instructions for use SMARTmatic S10 REF 1.011.6730 SMARTmatic S10 K REF 1.011.6731 SMARTmatic S10 S REF 1.011.6732 5 Operation | 5.1 Attaching the medical device 5 Operation 5.1 Attaching the medical device CAUTION Detachment of the medical device during treatment.
  • Page 14: Removing A Straight Or Contra-Angle Bur

    Instructions for use SMARTmatic S10 REF 1.011.6730 SMARTmatic S10 K REF 1.011.6731 SMARTmatic S10 S REF 1.011.6732 5 Operation | 5.4 Removing a straight or contra-angle bur WARNING Use of unauthorized burs. Risk of injury. ▶ Comply with the instructions for use and the intended use of the bur.
  • Page 15: Modification For Contra-Angle Burs

    Instructions for use SMARTmatic S10 REF 1.011.6730 SMARTmatic S10 K REF 1.011.6731 SMARTmatic S10 S REF 1.011.6732 5 Operation | 5.5 Modification for contra-angle burs ▶ Turn the taper sleeve back into its initial position. 5.5 Modification for contra-angle burs Note Using contra-angle burs, the handpiece must be converted.
  • Page 16: Checking For Malfunctions And Troubleshooting

    Instructions for use SMARTmatic S10 REF 1.011.6730 SMARTmatic S10 K REF 1.011.6731 SMARTmatic S10 S REF 1.011.6732 6 Checking for malfunctions and troubleshooting | 6.1 Check for malfunctions 6 Checking for malfunctions and troubleshooting 6.1 Check for malfunctions CAUTION Missing or damaged O-rings.
  • Page 17 Instructions for use SMARTmatic S10 REF 1.011.6730 SMARTmatic S10 K REF 1.011.6731 SMARTmatic S10 S REF 1.011.6732 6 Checking for malfunctions and troubleshooting | 6.2 Troubleshooting ▶ Use the nozzle needle (Mat. no. 0.410.0931) to free the water passage at the spray tubes.
  • Page 18: Processing Steps In Accordance With Iso 17664

    Instructions for use SMARTmatic S10 REF 1.011.6730 SMARTmatic S10 K REF 1.011.6731 SMARTmatic S10 S REF 1.011.6732 7 Processing steps in accordance with ISO 17664 | 7.1 Preparations at the site of use 7 Processing steps in accordance with ISO 17664 7.1 Preparations at the site of use...
  • Page 19: Cleaning

    Instructions for use SMARTmatic S10 REF 1.011.6730 SMARTmatic S10 K REF 1.011.6731 SMARTmatic S10 S REF 1.011.6732 7 Processing steps in accordance with ISO 17664 | 7.3 Cleaning ▶ Check the patency of the spray tube and clean it using the nozzle needle (Mat. no.
  • Page 20: Automated External Cleaning

    Instructions for use SMARTmatic S10 REF 1.011.6730 SMARTmatic S10 K REF 1.011.6731 SMARTmatic S10 S REF 1.011.6732 7 Processing steps in accordance with ISO 17664 | 7.4 Disinfection ▶ Brush under flowing tap water. 7.3.2 Automated external cleaning KaVo recommends washer disinfectors in compliance with ISO 15883-1, which are operated using alkaline cleaning agents.
  • Page 21: Manual External Disinfection

    Instructions for use SMARTmatic S10 REF 1.011.6730 SMARTmatic S10 K REF 1.011.6731 SMARTmatic S10 S REF 1.011.6732 7 Processing steps in accordance with ISO 17664 | 7.5 Drying 7.4.1 Manual external disinfection KaVo recommends the following products based on material compatibility. The micro- biological efficacy must be ensured by the disinfectant manufacturer and proven by an expert opinion.
  • Page 22: Servicing With Kavo Spray

    Instructions for use SMARTmatic S10 REF 1.011.6730 SMARTmatic S10 K REF 1.011.6731 SMARTmatic S10 S REF 1.011.6732 7 Processing steps in accordance with ISO 17664 | 7.6 Care products and systems - Servicing NOTICE Improper service and care. Premature wear and reduced product service life.
  • Page 23: Servicing With Kavo Quattrocare Plus

    Instructions for use SMARTmatic S10 REF 1.011.6730 SMARTmatic S10 K REF 1.011.6731 SMARTmatic S10 S REF 1.011.6732 7 Processing steps in accordance with ISO 17664 | 7.6 Care products and systems - Servicing See also: 2 Instructions for use KaVo QUATTROcare 2104 / 2104A ▶...
  • Page 24: Packaging

    Instructions for use SMARTmatic S10 REF 1.011.6730 SMARTmatic S10 K REF 1.011.6731 SMARTmatic S10 S REF 1.011.6732 7 Processing steps in accordance with ISO 17664 | 7.7 Packaging Note Close the chuck service mode. Option 1: Place the dental handpieces in the QUATTROcare PLUS 2124 A, close the front door and start the service procedure.
  • Page 25: Storage

    Instructions for use SMARTmatic S10 REF 1.011.6730 SMARTmatic S10 K REF 1.011.6731 SMARTmatic S10 S REF 1.011.6732 7 Processing steps in accordance with ISO 17664 | 7.9 Storage 7.9 Storage Processed products must be stored, protected from bacteria, to the extent possible, and dust, in a dry, dark, cool room.
  • Page 26 Instructions for use SMARTmatic S10 REF 1.011.6730 SMARTmatic S10 K REF 1.011.6731 SMARTmatic S10 S REF 1.011.6732 8 Tools and consumables 8 Tools and consumables Available from dental suppliers. Material summary Mat. No. Instrument stand 2151 0.411.9501 Cleanpac 10 units 0.411.9691...
  • Page 27 Instructions for use SMARTmatic S10 REF 1.011.6730 SMARTmatic S10 K REF 1.011.6731 SMARTmatic S10 S REF 1.011.6732 9 Terms and conditions of warranty 9 Terms and conditions of warranty The following Terms and conditions of warranty apply to this KaVo medical device:...

Table of Contents