5
Advanced Setup Functions

5.4 Setting the TDS Factor (Micro 600 TDS only)

The concentration of salts dissolved in solution increases the conductivity. This
relationship varies from salt to salt and is roughly linear over a given range for
a given salt. The TDS conversion factor is the number
used by the meter to convert from conductivity to TDS.
The TDS conversion factor can be set from 0.4 to 1.0.
Default value is 0.5.
See Section 13 - Calculating TDS Conversion Factor.
1 Enter advanced setup as described in Section 5.1.
2 Press s or t until [tdS] appears. Press ENTER.
3 Press s or t to select the desired TDS factor.
4 Press ENTER to select and return to the [tdS]
setup function.
5 Press s or t to move to the next setup function
or press CAL to exit to measurement mode.

5.5 Temperature Coefficient

The temperature coefficient is the amount of change in conductivity per degree
temperature (% per °C). For best results, determine and enter the exact
temperature coefficient of your solution. The meter allows adjustment from
0.0 to 3.0% per °C. Default value is 2.1% per °C.
Note: Select 0.0% for uncompensated
measurements. Temperature can be
measured by probe and displayed.
See Section 14 - Calculating
Temperature Coefficients.
1 Enter advanced setup as described in Section 5.1.
2 Press s or t until [t.Co %] appears. Press ENTER.
3 Press s or t to select a value between 0.0 to 3.0%.
4 Press ENTER key to select and return to the
[t.Co %] setup function.
5 Press s or t to move to the next setup