5
Advanced Setup Functions
5.8 Restore Factory Default Values
Use this function to reset all parameters to factory default settings. This clears
all calibration data and any other setup functions you might have changed.
IMPORTANT: Once activated the settings and
calibration data will be erased and can not be undone.
1 Enter advanced setup as described
in Section 5.1.
2 Press s or t key until [UrSt]
appears. Press ENTER.
3 Press s or t to select [Yes] or [no].
4 Press ENTER to confirm.
5 If reset [Yes] confirmed, the meter
will return to measurement mode
after initialisation.
22