Advanced Setup Functions

5.1 Advanced Setup Overview

Advanced setup allows customisation settings such as; selecting electrode's cell
constant, normalisation temperature, temperature coefficient, TDS factor (Micro
600 TDS), automatic or manual calibration (Micro 600 COND), single-point or
multi-point calibrations (COND and TDS), and to reset meter to factory default.
To enter advanced setup mode:
1 Switch off the meter.
2 Press MODE and then ON, holding both keys for 2 seconds.
Release the ON key before releasing the MODE key.
3 If successful, [StUP] will appear briefly followed by [CELC].
4 Press s or t to select the desired advanced setup function.
18
Entering Advanced Setup Mode
Cell Constant. Select k = 0.1, 1.0, or 10.0.
Default value is 1.0.
Select Automatic Calibration. "Yes" for auto calibration and "No"
for manual calibration. Default value is "Yes". (COND only)
Adjust Temperature Coefficient from 0.0 to 3.0% per °C.
Default value is 2.1% per °C.
Normalisation Temperature. Select "20°C" or "25°C".
Default value is 25 °C.
Adjust TDS factor from 0.4 to 1.0.
Default value is 0.5. (TDS only)
Select Single Point Calibration. Select "Yes" or "No".
Default value is "Yes".
User reset to factory defaults. Select "Yes" or "No".
Default value is "no".
Overview of Advanced Setup