5.2 Select Cell Constant

Your meter includes a probe with a nominal cell constant (k) of 1.0. Use probes
with k = 0.1 and 10 (sold separately) for improved performance in extreme samples.
Use this setup function to change the cell constant if necessary
Meter default is 1.0 to match the included probe.
k = 0.1 ideal for low measurements <20 µS/cm (<10 ppm).
k = 1.0 ideal for mid-range measurements
k = 10 ideal for high measurements >20 mS/cm (>10 ppt).
1 Enter advanced setup as described in Section 5.1.
2 Press s or t until [CELC] appears. Press ENTER.
3 Press s or t to select "1.0", "0.1" or "10.0".
Ensure that the cell constant corresponds with
the electrode you are using.
4 Press ENTER to select and return to [CELC]
setup function.
5 Press s or t to move to the next setup function
or press CAL to exit to measurement mode.

5.3 Automatic Calibration (Micro 600 COND only)

Select automatic calibration "YES" for easy
calibration of (4) factory pre-set conductivity
calibration standards (see Section 3.3 Table 1).
To use other standards or to calibrate any standard
manually select "no". Default value is "YES".
1 Enter advanced setup as described in Section 5.1.
2 Press s or t until [ACAL] appears. Press ENTER.
3 Press s or t to select [YES] or [no].
4 Press ENTER to select and return to
[ACAL] setup function.
5 Press s or t to move to the next setup function
or press CAL to exit to measurement mode.
Advanced Setup Functions