Switching The Meter On - palintest Micro 600 COND Manual

Conductivity/tds meters
Table of Contents

Advertisement

2
Getting Started
1 DO NOT measure or calibrate without
the protective probe guard in place.
2 Immersion above the protective guard
is not recommended. The cable can be
submerged briefly but is not designed
for continuous immersion.
See "Probe Care and Maintenance"
for more information.

2.8 Switching the Meter On

Press ON/OFF to power up your meter. Your meter will cycle through
various setup parameters when switched on.
1 The first screen shows the model [Con6] [tdS6].
2 The second screen shows the nominal cell constant value. The Micro 600
COND and TDS meters can accept electrodes with k = 0.1, 1.0 or 10.0
nominal cell constants. Default value is k = 1.0 [C 1.0]. See Section 5.2
Advanced Setup to modify.
3 The third screen shows the Normalisation Temperature which can be set
to 25°C or 20°C. Default value is 25°C [t 25.0°C]. See Section 5.6
Advanced Setup to modify.
4 The fourth screen shows the Temperature Coefficient which can be set from
0.0 to 3.0% per °C. Default value is 2.1%/°C [t 2.1%]. See Section 5.5
Advanced Setup to modify.
5 All LCD segments light for 2 seconds before entering measurement mode.
Note: the meter will use the measurement mode that was in
use when it was previously powered off.
8
Immerse
probe
beyond
upper
steel
band