Select the Hotkey Test List from the Mode menu. The ten
available positions are listed along with assigned or empty
slots.
Select the Hotkey to assign (0 – 9) and use the up/down
arrows to highlight the required parameter/method from the
Choose a Test list.
Press OK and the test will be assigned to the defined number.
5.7 Units
The Photometer 7500 Bluetooth offers the choice of result
.
expressed in mg/l, ppm, mmol/l, µmol, g/l and µg/l
Changing the result units will not affect the result log.
5.8 Dilution Factor
When samples are above the test range, indicated by >> on
the result screen, a dilution procedure can be used. Setting
Dilution Factor to On will prompt the Photometer 7500
Bluetooth to automatically request the dilution factor when
carrying out a test.
Change the dilution factor by using the up/down arrows or
manually type the dilution factor. The instrument will
automatically correct the result for the dilution and display the
corrected result (which will also be stored in the result log).
The maximum dilution factor permitted is 99.
If the calculated result exceeds the available number of
permitted characters >> will be displayed
23