4.4 Using Manual Ranging Function

By default your meter has auto-ranging ability and automatically selects the
range in which your readings appear.
Alternatively, to override auto-ranging you can manually select a specific range
by pressing s successively for each measurement range. The five ranges are:
Micro 600 COND
Conductivity Range
0 - 20.00 µS/cm
0 - 200.0 µS/cm
0 - 2000 µS/cm
0 - 20.00 mS/cm
0 - 200.0 mS/cm
Auto ranging:
Manual ranging:
0 - 200.0 mS/cm
NOTE: If the value of the solution you are measuring is higher than the range
selected [Or] (over range) will appear. Press s to select a measurable range.
The meter resets to Auto-ranging function once it is turned off. You will
have reset the manual ranging function each time you turn the meter off.
16
Micro 600 TDS
TDS Range (using 0.5 TDS factor)
0 - 10.00 ppm
0 - 100.0 ppm
0 - 1000 ppm
0 - 10.00 ppt
0 - 100 ppt
Manual ranging:
0 - 20.00 uS/cm
Manual ranging:
0 - 20.00 mS/cm
Manual ranging:
0 - 200.0 uS/cm
Manual ranging:
0 - 2000 uS/cm