Cleaning, Disinfection and Sterilization
7.5.2

Machine-based Reprocessing of the Contra-Angle

(Thermal-Disinfection)
Operation
Automatic cleaning /

disinfection

Cleaning at 65°C
(149°F) holding time
A
5 min.
Disinfection at 90°C
(194°F) holding time
5 min.
Ensure the contra-angle attachment is completely dry after
thermal disinfection.
Remove any residual liquid with particle-free compressed air.
Drying is performed automatically during the drying cycle in
the washer-disinfector.
Sterilization is required after Machine-based Reprocessing of
the Contra-angle.
T US XSIQ DFU WEB / Rev.00 (based from T US XSIQ DFU WEB / Rev.05 / 07-2017) / 09-2018 65/88
Use washer-disinfector conforming to EN
ISO 15883, e.g. Belimed WD, programme:
neutral.
Use cleaning solution: e.g. 0.5% (v/v)
deconex 24 LIQ, Borer Chemie.
Use neutralizing solution: e.g. 0.25% (v/v)
deconex 26 plus, Borer Chemie.
Deionized water (DI) or water that has this
degree of purity is recommended for
cleaning.
Remarks