GasAlertMicro 5/PID/IR
User Manual
Applying Gas to the Sensors
The calibration cap, single gas calibration cap, and hose are shipped
with the detector. Refer to
Figure 3.
The calibration cap and single gas calibration cap must only be
used during the calibration span process.
Figure 3. Applying Gas to the Sensors
50
Find Quality Products Online at:
and
Table 11.
for installation.
Note
GlobalTestSupply
www.
Table 11. Applying Gas to the Sensors
Item
1
Detector with calibration cap
2
Calibration hose
3
Regulator and gas cylinder
4
Toxic 2 sensor position
5
Single gas calibration cap
Single Gas Calibration Cap
a Caution
If an O
or ClO
sensor is located in the Toxic 2 position
3
2
(refer to
Figure 3.
and
Table
cap must be used to ensure accurate calibration.
To calibrate O
and ClO
sensors using the single gas calibration cap,
3
2
refer to
Figure
4.,
Table
12., and complete the following procedures.
1. Insert the cap into the Toxic 2 sensor position on the detector
(refer to
Figure
3.). Press firmly until the release tabs click.
2. Connect the calibration hose to the gas cylinder and to the
intake inlet on the cap.
3. Proceed to
Calibration
Procedure.