Honeywell BW Ultra User Manual
Honeywell BW Ultra User Manual

Honeywell BW Ultra User Manual

Portable five-gas detector with internal pump
Hide thumbs Also See for BW Ultra:

Advertisement

Quick Links

User Manual
| Gas Detection
Honeywell BW™ Ultra
Portable Five-gas Detector
with Internal Pump

Advertisement

Table of Contents
loading

Summary of Contents for Honeywell BW Ultra

  • Page 1 User Manual | Gas Detection Honeywell BW™ Ultra Portable Five-gas Detector with Internal Pump...
  • Page 2 Honeywell until receipt by the buyer of shipping instructions from Honeywell. • The right of Honeywell to require that the buyer provides proof of purchase such as the original invoice, bill of sale, or packing slip to establish that the product is within the warranty period.
  • Page 3: Table Of Contents

    Sensor T90 Ratings.........................17 Activate the backlight ........................ 7 Label Information........................18 Deactivate the detector ......................7 BW Ultra Battery label ....................18 BW Ultra Instrument label ..................20 Navigate the menu ........................7 Part Number Format .......................22 View detector’s general information ................... 7 Sensor Replacement .......................23...
  • Page 4: Before You Begin

    Oxygen (O2) % volume In no event is Honeywell liable to anyone for any indirect, special, Combustible gases (LEL) a) percent of lower explosive limit (%LEL) or consequential damages. The information and specifications Lower explosive limit b) percent by volume methane 0-5.0% v/v...
  • Page 5: Safety Information

    • The lithium battery in this product presents a risk of fire, explosion, and chemical • DO NOT rely on the Bluetooth BW Ultra output indication for security purposes. burn if misused. Do not open, crush, disassemble, incinerate, or heat above 100°C (212°F).
  • Page 6: Sensor Poisons And Contaminants

    Electrical Equipment for Explosive Atmospheres agréé ou un centre de traitement des matières dangereuses. Éloignez les enfants des piles au lithium. • Ne comptez PAS sur l’indication de sortie Bluetooth BW Ultra à des fins de Natural Institute of Metrology, Quality, and sécurité.
  • Page 7: Getting Started

    Bump test failed TWA alarm Correction factor Peak gas exposure Firmware update in Over limit alarm progress High alarm Sensor disabled This icon flashes to indicate regular Low alarm operation without any gas alarms or functional errors BW Ultra User Manual...
  • Page 8: Alarms

    Depending on the type of Fast flash alarm, these flashes, vibrations, and noises will be different. NOTE: In Stealth mode the Honeywell BW™ Ultra only Black box around gas Over Limit vibrates. flashes...
  • Page 9: Alternate Display Background

    Alternate Display Background To activate its functionality: 1. Go to Instrument Operation Tab. 2. Check or uncheck the Normal Display box to activate or deactivate the White Background. BW Ultra User Manual...
  • Page 10: Detector Operations

    Configurator to reset readings in the detector. 1. Go to the main menu and select > See Information > Peak readings. 2. Select Hold to reset peak readings. Press and hold for 3 seconds to reset readings. BW Ultra User Manual...
  • Page 11: Reset Twa/Stel And Peak Readings

    Passed will be displayed. Press the button to exit be bypassed. calibration. If the calibration failed for some or all of the gases, either a Cal Error All gases applied mixed results message (if the detector was not successfully BW Ultra User Manual...
  • Page 12: Select Pid Correction Factor

    15 minutes between calibrations and all the nests used in gang should be for BW Ultra. If it is not the case, it 9. The detector will prompt you Did you see and hear...
  • Page 13: Set Up Reverse Intelliflash

    The detector can be used in one of three modes: Essentials mode, Hole Watch mode, and Inert mode. : Honeywell BW™ Ultra always monitors gas levels, regardless of Note the operating mode. If the detector detects a sudden gas exposure, it will flash, vibrate, and produce a loud siren noise.
  • Page 14: Configure The Detector

    General Sensor Settings (for all sensors) • Sensor Enable/Disable • Calibration Gas Concentration : When the operator configures the Honeywell BW™ Ultra using Note Safety Suite Device Configurator, Honeywell strongly recommends reviewing the detector’s settings before the operation to ensure that settings were applied successfully and comply with performance requirements.
  • Page 15: Maintenance

    The type of sensor can only be configured at within 8 hours. factory during assembly process. Due to above, the user request a BW Ultra with a dummy sensor in the Analog sensor position, then this dummy Maintain the Battery position will not be able to configure and install an analog sensor.
  • Page 16: Languages

    Languages button to see the connected devices. 3. Click over the device serial number. Honeywell BW™ Ultra supports six languages: English, 4. Click the Update button on the Device Details screen French, German, Portuguese, Spanish, Simplified Chinese. below the Firmware Version section.
  • Page 17: Appendices

    10 ppm 15 ppm 35 ppm 50 ppm 35 ppm 200 ppm 19.5% vol. 23.5% vol. 10% LEL 20% LEL Note: To disable an alarm, set the alarm setpoint to 0 (zero) in Safety Suite Device Configurator.. BW Ultra User Manual...
  • Page 18: Standards And Certifications

    II 1 G Ex da ia IIC T4 Ga, -40°C ≤ Tamb ≤ +50°C (with LEL sensor installed and without IR sensor installed) The Honeywell BW™ Ultra gas detector is in conformity with the following standards and certifications: I M1 Ex db ia I Ma, -20°C ≤ Tamb ≤ +50°C II 2 G Ex db ia IIC T4 Gb, -20°C ≤...
  • Page 19 IP54. Other IP ratings are not the scope of ISA 60079- This device complies with FCC radiation exposure limits 29-1. Honeywell BW™ Ultra was pressure tested for 80, 100 set forth for an uncontrolled environment and meets the and 120 kPa in ISA 60079-29-1. Outside of 80 - 120 kPa is FCC radio frequency (RF) Exposure Guidelines.
  • Page 20: Sensor T90 Ratings

    Sensor T90 Ratings Please be advised of the following T90 ratings for the BW Ultra sensors stated below: Sensor Type T90 Ratings LEL Sensor (1LEL75C) for Methane is < 20 sec @ 20°C Oxygen (O2) Sensor: 1O2 for O2 is typically <15 sec Carbon Monoxide (CO) Sensor: CO 4R+ for CO is <...
  • Page 21: Label Information

    Label Information BW Ultra Battery label cULus ATEX/UKEX IECEx JPEx 1. BW Ultra battery serial number and2D barcode 2. Manufacturer information 3. CEC battery charging system approval marking 4. WEEE marking 5. Chinese EPUP marking cULus BW Ultra User Manual...
  • Page 22 ATEX/UKEX IECEx BW Ultra User Manual...
  • Page 23: Bw Ultra Instrument Label

    4. Class Zone protection markingwhen BW Ultra is configured with LEL and without IR sensor 5. Class Zone protection marking when BW Ultra is configured with IR sensor and without LEL sensor 6. Class Zone protection marking when BW Ultra is configured with IR sensor and with LEL sensor 7.
  • Page 24 XXXX XXXX UL21UKEX2006X UL21UKEX2006X UL21UKEX2006X 5. ATEX certificate number 6. BW Ultra part number and 2D barcode 7. BW Ultra serial number and 2D barcode IECEx Ex ia I Ma, Ex ia I Ma, Ex ia I Ma, Ex ia IIC T4 Ga...
  • Page 25: Part Number Format

    Part Number Format BW Ultra’s explosion-proof protection type varies per the sensor configuration. The user can notice the sensor configuration with the part number format. B. (5 mm X 5 mm) 2D barcode for Part Number: 14 digits 1 digit for country code...
  • Page 26: Sensor Replacement

    4HYT, manufactured by City 50122982-089 032-0121-000, manufactured by RAE Systems 50122982-090 032-0112-000, manufactured by RAE Systems 50122982-091 4HN, manufactured by City 50122982-092 032-0111-000, manufactured by City 50122982-210 CO-AX, manufactured by Alphasense 50122982-203 Empty sensor, manufactured by RAE Systems BW Ultra User Manual...
  • Page 27: Troubleshooting

    Detector automatically deactivates. Automatic deactivation due to critical low Replace the battery. battery. Lockout on Self-Test Error is enabled and Change the sensor. a sensor(s) has failed the startup self-test. Sensor(s) require calibration. Calibrate the detector. BW Ultra User Manual...
  • Page 28: Glossary

    Safety Suite Device Configurator accumulated averages of gases. Using the US Occupational Safety and A proprietary, Windows-based software developed by Honeywell to Health Administration (OSHA) method or the American Conference of configure and manage docking modules, calibration, bump tests, and data Governmental Hygienists (ACGIH) method, an average is calculated to logs.
  • Page 29: Contact Honeywell

    Honeywell Analytics 405 Barclay Boulevard Lincolnshire, Illinois 0069 USA 60069 Toll free: 1-888-749-8878 Asia Honeywell Analytics Asia Pacific 7F SangAm IT Tower, 434 Worldcup Buk-ro, Mapo-gu, Seoul 03922, Republic of Korea Tel: +82 (0) 2 6909 0300 Analytics.ap@honeywell.com Europe Honeywell Analytics...

Table of Contents