Setting the Calibration Due Date
7. When the span is complete, the calibration due date can be set
for each sensor that has spanned successfully. The following
screen displays.
Press C to set the calibration due dates.
Or
Press A to bypass and proceed to
If a new due date is not entered, the detector automatically
resets to the previously entered number of days.
Find Quality Products Online at:
Alarm Setpoints
step #8.
Note
GlobalTestSupply
www.
Calibration and Setting Alarm Setpoints
The calibration due dates are set in the following order:
• Toxic 1
• Toxic 2
• LEL
• O
2
If the due date of an unsuccessfully spanned sensor is
changed, the following screens display.
The detector then automatically proceeds to the next sensor.
To change the calibration due date (1-365 days), press H or
G until the new value displays. Press C within 5 seconds to
confirm.
Or
Press A to bypass a sensor and proceed to the next sensor.