Time Required To Span - Honeywell BW GasAlertMicro 5 User Manual

Table of Contents

Advertisement

Attach the calibration cap (or single gas calibration cap for O
and ClO
) and apply gas to the sensor(s). To attach caps, refer
2
to
Figure
3..
Refer to the following flow rates:
• 1000 ml/min. for NH
• 500 ml/min. for CO
• 250-500 ml/min. (all other sensors)
K flashes as the detector initially detects the calibration gas.
GasAlertMicro 5 and PID
After 30 seconds the detector beeps and K stops flashing.
flashes while spanning the sensors until the detector
has attained a sufficient level of the expected gas.
Refer to
Table 13.
for times required to span.
Table 13. Time Required to Span
Gas Type
Most toxic gases and CO
Find Quality Products Online at:
and Cl
3
2
2
GasAlertMicro 5 IR

Time Required to Span

2 minutes
2
GlobalTestSupply
www.
3
Exotic toxic gases
LEL (combustibles)
PID gases
Insufficient Level: If a sensor does not attain a sufficient level
of expected gas, it is cleared from the LCD and is not spanned.
.com
GasAlertMicro 5/PID/IR
Calibration and Setting Alarm Setpoints
5 minutes
30 seconds
2 minutes
sales@GlobalTestSupply.com
55