Confined Space Sampling

The sample probe is used to safely test for gas in confined spaces
before entering.
Attach the sample probe to the pump module prior to activating the
detector.
To measure hazardous gas in a confined space, the
sample probe must be used with the pump module.
70
Find Quality Products Online at:
a Warning
GlobalTestSupply
www.
Figure 11. Attaching the Sample Probe
a Caution
To prevent the Teflon lining inside the Tygon tubing from
causing a blockage when connecting it to the sample
probe, the end of the tubing must be flared. Refer to
steps 3 and 4.
Note
The detector and pump module can also be used with a sin-
tered filter.
1. Deactivate the detector.
2. Attach the pump connector end of the teflon-lined Tygon tubing
to the nozzle on the pump module.
3. Gently insert needle nose pliers into the other end of the teflon-
lined Tygon tubing. Using a circular motion, flare the end of the
tubing.
4. Connect the flared end of the tubing to the sample probe.
Ensure the teflon lining does not separate from the Tygon tub-
ing, as it will block the tube and generate a pump alarm.
5. Activate the detector. Ensure all connections are secure before
sampling.
6. Insert the sample probe into the confined space.
a Warning
Depending upon the length of the tubing and the type of
gas in the confined space, allow a minimum of 3
.com