Initialize

The Initialize option restores the origi-
nal factory default settings of the detec-
tor.
1.
Press H to scroll to Initialize and
press C within 20 seconds to
confirm.
2.
From the Initialize? screen
perform one of the following options
within 5 seconds.
• Press A No to exit, or
• Press C Yes to initialize.
If No is selected, the following screen
displays and the detector exits the initial-
ize option.
Find Quality Products Online at:

Force Calibration

The detector is shipped with the Force cal option disabled.
GlobalTestSupply
www.
If Yes is selected, the following screen
displays while performing the initializing
process.
When initializing is complete, the
following screen displays.
The detector deactivates and then
immediately reactivates. The detector
then reconfigures to the default settings
while it performs the self-test.
Verify all settings and alarm setpoints,
and then calibrate the sensors.
If enabled, the Force cal option
automatically forces the detector to enter
calibration if a sensor is overdue upon
startup. Press H to scroll to Force cal.
Press C to toggle between enable and
disable.