Honeywell BW GasAlertMicro 5 User Manual page 69

Table of Contents

Advertisement

If a new setpoint is entered but not confirmed within 5 seconds
by pressing C, the following screen displays.
The previous setpoint is retained and the detector
proceeds to the next setpoint.
The setpoints are set in the following order:
• TWA (if applicable)
• STEL (if applicable)
• low
• high
To bypass a setpoint, press C to save the current value and
proceed to the next setpoint.
Find Quality Products Online at:
Setting the TWA Alarm Setpoint
The current TWA alarm setpoint displays for the selected sensor
(if applicable).
Press H or G to change the value for the TWA alarm setpoint. When
the required value displays, press C to confirm.
Setting the STEL Alarm Setpoint
The current STEL alarm setpoint displays for the selected sensor
(if applicable).
Press H or G to change the value for the STEL alarm setpoint. When
the required value displays, press C to confirm.