GasAlertMicro 5/PID/IR
User Manual
Problem
Detector Operation Troubleshooting
Detector does not display
normal ambient gas
readings after startup
self-test.
Detector does not respond
to buttons.
Detector does not
accurately measure gas.
90
Find Quality Products Online at:
Possible Cause
Sensor not stabilized
Detector requires calibration
Target gas is present
Batteries are depleted
Detector is performing
operations that do not require
user input
Detector requires calibration
Detector is colder/hotter than
ambient gas
Sensor filter is blocked
GlobalTestSupply
www.
Used sensor: wait 60 seconds
New sensor: wait 5 minutes
Calibrate the sensors. Refer to
Detector is operating properly. Use caution in suspect areas.
Refer to
Replacing the Alkaline Batteries
Pack.
Button function restores automatically when the operation ends.
Calibrate the sensors. Refer to
Allow the detector to adjust to ambient temperature before using.
Clean the sensor filter. Refer to
.com
Solution
Calibration and Setting Alarm
Setpoints.
or
Replacing the Lithium Battery
Calibration and Setting Alarm
Setpoints.
Replacing a Sensor or Sensor
sales@GlobalTestSupply.com
Filter.