System Controller Kit; General Equipment Information - Zimmer Biomet iASSIST Knee Manual

2-pod version, surgical technique
Table of Contents

Advertisement

System Controller Kit

Before cleaning the system controller computer,
always unplug computer from the AC power outlet to
prevent electrical shocks.
Use ready to use surface disinfectant wipes (for
example Accel™ TB Wipes) for cleaning/disinfection.
Carefully wipe all surfaces of the system controller
computer including the touch screen if necessary.
Hardware
System Controller Computer
System Controller Power Supply
System Controller Carrying Case
System Controller Shipping Case

General Equipment Information

Warning: If one of the following situations arises,
have the equipment verified by a qualified service
personnel:
a. The equipment is not functioning properly nor
functioning as per instructions;
b. The system controller computer or the
instruments have obvious/visible signs of
breakage or fatigue;
c. The system controller computer or the
instruments have been dropped and are
damaged;
d. Liquid has entered the system controller
computer;
e. The power cord or plug is damaged; or
f. If the date and time function of the system
controller computer stops functioning properly.
If an instrument marked with the below symbol
is dropped, have the instrument verified by a
qualified service personnel.
Do not use aerosols, solvents, strong detergents,
or corrosive products. Never allow water or other
liquids to enter the computer since this may cause
subsequent short-circuits or corrosion.
Product ID
20-8011-070-01
20-8011-070-02
System Controller Carrying Case
System Controller Shipping Case
System Controller Kit
The system controller computer has an internal
battery that allows three hours of autonomy. This
power backup is provided to prevent data loss if
there is a main power outage.
If the system controller computer needs to be
shipped with a carrier, place the system controller
carrying case in the system controller shipping
case. The shipping case is specifically designed to
withstand carrier shipping conditions.
The expected life of the computer under normal
usage conditions is five years.
iASSIST 2-Pod Version Surgical Technique
51 |
Warning: Disinfecting computer equipment with
sprays is not recommended since the vapor can
enter the equipment which may cause electrical
short-circuits or corrosion.
Qty
Cleaning Instructions
1
See above
1
See above
1
See above
1
See above
Note: When these instruments are dropped,
their precision may be impaired. The mechanical
deformity may be difficult to detect visually.