Essential Performance
The iASSIST Pod contains inertial sensors (accelerometers and gyroscopes) and instrument identification switches.
The following performances have been identified as essential to the iASSIST Pod:
1. The accelerometer inclination measurements on all axes are stable within ±1.0°.
2. The gyroscope rotation speed measurements on all axes are stable within ±1.0°/sec.
3. The instrument identification switches detect the right instrument code.
These essential performances have been tested during electromagnetic immunity tests up to the levels shown in
table 5.6 and 5.7. Loss or degradation of essential performance can result in navigation errors and/or instrument
misidentification.
Guidance and Manufacturer's Declarations
Table 5.5: Guidance and Manufacturer's Declaration – Electromagnetic Emissions
The pods are intended for use in the electromagnetic environment specified below.
The customer or the user of the pods should assure that it is used in such an environment.
Emissions test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emissions
Not applicable
IEC 61000-3-2
Voltage fluctuations/
Not applicable
flicker emissions
IEC 61000-3-3
Electromagnetic environment – guidance
The pods use RF energy only for its internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby electronic equipment.
The pods are suitable for use in all establishments, including domestic establishments and
those directly connected to the public low-voltage power supply network that supplies
building used for domestic purposes.