Zimmer Biomet iASSIST Knee Manual page 20

2-pod version, surgical technique
Table of Contents

Advertisement

iASSIST 2-Pod Version Surgical Technique
18 |
Distal Femoral Cut Validation
1. Remove the pod attached to the femoral reference
and clip it to the validation tool.
2. Position the flat surface of the validation tool onto
the distal femoral cut.
3. Secure the validation tool on the distal femoral
cut using the captive spikes. Additionally two
3.5 X 38 mm hex head screw can be used for more
stability.
Warning: Care must be taken when impacting
the spikes of the validation tool to avoid
potential interference between the spikes and
the pins used to secure the femoral distal cut
guide and the femoral saw slot. If interference
is felt, the validation tool should be shifted
mediolaterally by 10 mm.
Figure 3.6
Distal Femur Cut – Validation
For steps four through six, follow the feedback on the
pod attached to the validation tool. Ensure adequate
stability of the validation tool on the distal femoral cut
by manually holding it in place.
4. Bring the leg into abduction. Once the green
LED stops blinking hold the knee steady until the
acquisition sound is heard.
5. Bring the leg into adduction. Once the green
LED stops blinking hold the knee steady until the
acquisition sound is heard.
6. Bring the leg into neutral position. Once the green
LED stops blinking hold the knee steady until the
acquisition sound is heard.
7. The distal femur cut values are displayed on the
pod attached to the distal femoral cut guide.

Advertisement

Table of Contents
loading

Table of Contents