Distal Femoral Cut (cont.)
5. Once the adjustment is complete, slide the femoral
adjustment mechanism towards the femur until it
is fully seated on the most distal condyle.
6. Insert three 3.2 mm headless trocar drill pins in
the base of the distal femoral cut guide using both
parallel holes and one diagonal hole in the body of
the guide.
Figure 3.5
Distal Femur Cut
7. With the knee stable and flexed between 45 and
80 degrees, and the leg in neutral abduction,
unlock the femoral A/P slider using the blue screw
and remove it from the assembly. Hold the knee
steady until the confirmation sound is heard.