1
4
Powering-On Pods and
Intraoperative Calibration (cont.)
3. Wait for the green status LED of all of the pods to
start blinking slowly before continuing with the
intraoperative calibration.
Warning: The calibration process, including
the pod calibration jig handling, should be
performed wearing sterile attire.
4. Clip the pods to the pod calibration jig. When
successfully clipped, the green status LED on the
pods will turn solid green.
2
5
Figure 3.1
Powering-On Pods and intraoperative Calibration (Continued)
iASSIST 2-Pod Version Surgical Technique
9 |
3
5. Begin by placing the first face of the pod calibration
jig upwards on a sterile table. Keep the jig steady
until a sound is triggered. Continue the sequence
by positioning the jig in its remaining five positions
(2 through 6). When the calibration is completed,
the green status LED of each pod will blink.
Note: The green LEDs on the flexion/extension
axis of the cut guide pod indicate the expected
position (1 through 6) of the pod calibration jig
during calibration.
Warning: The pods need to be calibrated on a
flat, approximately horizontal surface to ensure
optimal performance. This surface must be
sterile and stable during the manipulations.
6. Unclip all pods from the pod calibration jig.
6