Table 5.4: Recommended separation distances between portable and mobile RF communications equipment
and the system controller computer
The system controller computer is intended for use in the electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the system controller computer can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the system
controller computer as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE
1
: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE
: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
2
structures, objects and people.
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d=1.2 √P
0.12
0.38
1.2
3.8
12
iASSIST 2-Pod Version Surgical Technique
59 |
m
80 MHz to 800 MHz
800 MHz to 2.5 GHz
d=1.2 √P
0.12
0.38
1.2
3.8
12
d=2.3 √P
0.23
0.73
2.3
7.3
23