Zimmer Biomet iASSIST Knee Manual page 27

2-pod version, surgical technique
Table of Contents

Advertisement

Proximal Tibial Cut
1. Unclip the cut guide pod from the tibial alignment
guide and clip it to the front receptable of the tibial
adjustment mechanism.
2. Follow feedback (red and green LEDs) on the pod
attached to the tibial adjustment mechanism and
adjust flexion/extension and varus/valgus using
the gold and green screws respectively.
Warning: The knee should be flexed between
45 and 80 degrees in order for the system to be
able to compute angle values.
When using the 3.5 mm hex head knob, remove
the knob before performing the cut.
Figure 3.9
Proximal Tibial Cut
iASSIST 2-Pod Version Surgical Technique
25 |