Condenser-side Emboss Contrast Slider
The condenser-side emboss contrast slider is equipped
with a sector diaphragm. Attaching a centering telescope
to the eyepiece tube enables you to view a sector
diaphragm image.
You can change the direction of image contrast by rotating
the condenser-side emboss contrast slider adjuster to turn
the sector diaphragm.
To use the condenser-side emboss contrast slider, first
remove the phase contrast slider from the condenser.
Then insert the condenser-side emboss contrast slider into
the condenser slider slot ①.
Eyetube-side Emboss Contrast Slider
The eyepiece-tube-side emboss contrast slider has several
position markings corresponding to the objective
magnification, and several stop positions to ensure
alignment of the apertures with the light path. For emboss
contrast microscopy, insert the slider into the microscope
until it reaches the position of the same number as the
magnification of the objective. To switch back to bright-
field microscopy, pull out the slider up to the hollow
position. Slider position ❶ corresponds with aperture ①,
❷ with ②, and so forth.
For observation without emboss contrast, ensure the
condenser-side emboss contrast slider is in the open
position, and the eyetube-side slider is in position ❶.
Using a Microscopy Camera
Installing Couplers (Refer to Page 16)
Selecting the Light Path for Observation/Imaging with a Camera (Refer to Page 21)
ACCU-SCOPE
®
73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com
EXI-410 INVERTED MICROSCOPE SERIES
(Optional)
Open
position
② ③ ④
Sector
diaphragm
Adjuster