Accu-Scope EXC-120 SERIES Manual
Hide thumbs Also See for EXC-120 SERIES:

Advertisement

Quick Links

MANUAL
EXC-120
MICROSCOPE SERIES
73 Mall Drive, Commack, NY 11725 • 631-864-1000 (P) • 631-543-8900 (F)
www.accu-scope.com • info@accu-scope.com

Advertisement

Table of Contents
loading

Summary of Contents for Accu-Scope EXC-120 SERIES

  • Page 1 MANUAL EXC-120 MICROSCOPE SERIES 73 Mall Drive, Commack, NY 11725 • 631-864-1000 (P) • 631-543-8900 (F) www.accu-scope.com • info@accu-scope.com...
  • Page 3: Table Of Contents

    ........................11 FOCUSING CONDENSER ......................11-12 INTERPUPILLARY DISTANCE ..................12 TENSION ........................13 FUSE REPLACEMENT ....................14 TROUBLESHOOTING ....................15-17 MAINTENANCE........................18 SERVICE ..........................18 WARRANTY ........................18 ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
  • Page 4: Safety Notes

    An annual schedule of preventative maintenance by qualified personnel is highly recommended. Your ® authorized ACCU-SCOPE distributor can arrange for this service. ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
  • Page 5: Introduction

    Carefully remove the microscope from the Styrofoam container by its arm and base and place the microscope on a flat, vibration-free surface. ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
  • Page 6: Components Diagrams

    Eyepiece Nosepiece Y Axis Objective Adjustment Knob Stage X Axis Adjustment Knob Condenser Fine Focusing & Filter Holder Knob Illuminator Coarse Tension Focusing Adjustment Knob Power Collar Switch ® 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com ACCU-SCOPE...
  • Page 7: Components Diagrams

    EXC-120 MICROSCOPE SERIES COMPONENTS DIAGRAM Interpupillary Distance Indicator Condenser Adjustment Knob Specimen Holder Illuminator Adjustment Dial ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
  • Page 8: Assembly Diagram

    When assembling the microscope, make sure that all parts are free of dust and dirt, and avoid scratching any parts or touching glass surfaces. Eyepieces ② Binocular Viewing Head ③ ① Objectives ④ ④ Main Body Power Cord ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
  • Page 9: Detailed Assembly

    Lock the eyepiece into place by tightening the lock screw (small hex screw) on the eyetube. Repeat above for the other eyepiece. ② ① Fig. 4 Fig. 3 ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
  • Page 10: Installing The Objectives

    ③ on the filter holder. Place the filter ④ into the top of the holder. ③ Swing the filter holder ③ back into place under the ① condenser. Fig. 7 ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
  • Page 11 Should you need a Fig. 8 replacement, contact your authorized ACCU-SCOPE dealer or call ACCU-SCOPE at 1-631-864-1000 for a dealer nearest you. ③ Fig. 9 ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
  • Page 12: Adjustment & Operation

    - forward/back (Y), and bottom knob - left/right Fig. 11 (X ). ④ ⑤ Fig. 12 ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
  • Page 13: Focusing

    Fig. 14 objective. Focus the specimen with the 10x objective. Adjust the condenser focusing knob ① to get a clear image of the field iris diaphragm. ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
  • Page 14: Interpupillary Distance

    Where “·” on the eyetube ① lines up, then that is the number for the interpupillary distance. Range:48~75mm (binocular models) 55~75mm (trinocular models) Remember your interpupillary for future operation. ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
  • Page 15: Tension

    Fig. 17 To increase tension, turn the tension adjustment collar counterclockwise to tighten; turn clockwise to loosen the tension. ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
  • Page 16: Fuse Replacement

    Specification of the fuse: 250V, 0.5A. Fig. 20 CAT # 3277 Fig. 21 Fig. 22 ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
  • Page 17: Troubleshooting

    Air bubbles in immersion oil Remove bubbles Condenser aperture is closed Open or close properly or open too much Condenser is positioned too Position the condenser slightly lower than the upper limit ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
  • Page 18 Cover glass is to thick Use a 0.17mm cover glass Diopter adjustment is not set Readjust the diopter settings properly as outlined in section 4.3 ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
  • Page 19 Increase the tension on the the coarse focusing knob focusing knobs Fine focus is ineffective Tension adjustment is set too Loosen the tension on the high focusing knobs ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
  • Page 20: Maintenance

    SCOPE at (631) 864-1000. This warranty is limited to the continental United States of America. All items returned for warranty repair must be sent freight prepaid and insured to ACCU-SCOPE INC., 73 Mall Drive, Commack, NY 11725 – USA. All warranty repairs will be returned freight prepaid to any destination within the continental United States of America, for all foreign warranty repairs return freight charges are the responsibility of the individual/company who returned the merchandise for repair.