Mounting A Microscopy Camera (Optional) - Accu-Scope EXI-310 Series Manual

Inverted microscope
Hide thumbs Also See for EXI-310 Series:
Table of Contents

Advertisement

Operation
(continued)
Centering the Light Annulus
NOTES:
Halo-like ghost images of the light annulus may sometimes appear. It this occurs, superimpose the
brightest light annulus image over the phase annulus.
When a thick specimen is moved or replaced, the light annulus and the phase annulus may deviate. This
is usually due to the amount of media or wellplate inconsistences. This can reduce image contrast. If this
occurs, repeat steps 1-5 for readjustment.
The centering procedure may have to be repeated in order to get the best possible contrast if a specimen
slide or the bottom surface of a culture vessel is not flat. Center the light annulus using objectives in the
order of lower to higher magnifications.
Using a Microscopy Camera
(Optional)
Installing Couplers
1. Remove the dust cap on the top photo port.
2. Attach the camera to a c-mount adapter.
3. Thread the c-mount adapter with camera in the top photo port.
Selecting the Light Path for Observation
With a Camera
Refer to page 12.
®
ACCU-SCOPE
EXI-310 INVERTED MICROSCOPE SERIES
(Continued)
73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com
15