Objectives
To install the objectives:
1. Turn the coarse adjustment knob ① until the revolving
nosepiece is at its lowest position.
2. Remove the nosepiece cap ② closest to you and
thread the lowest magnification objective onto the
nosepiece opening, then rotate the nosepiece
clockwise and thread the other objectives from low to
high magnification.
NOTE:
● Always rotate the nosepiece by using the
knurled nosepiece ring.
● Keep the covers on any unused
nosepiece openings to prevent dust and
dirt from getting inside.
Stage Plate
Insert the clear glass stage plate ① into the opening on
the stage. The clear glass allows you to view the
objective in position.
Eyepieces
Remove the eyetube plugs and fully insert the eyepieces
① into the eyepiece tubes ②.