Selector E makes it possible to configure the supply voltage of the device in a 115/230 V AC version.
→
Energize the device with a
230 V AC voltage.
Figure 12:
Selector of the supply voltage on a 115/230 V AC version
→
Loosen the nuts of the cable glands.
→
Insert each cable through a nut than through the cable gland, using the cable glands as shown in "Figure 13".
Flow sensor cable
Figure 13:
Using the cable glands
→
Remove the two terminal blocks (marked 6 and 7 in "Figure 10") from the housing.
Figure 14:
Inserting the cable clips
→
Depending on the operating voltage of the device, wire according to chap. "7.4.5" to "7.4.10".
→
Insert the two terminal blocks (marked 6 and 7 in "Figure 10") into their original position.
→
Letting the housing stay completely open, secure the power supply cable, the flow sensor connection cable
and, depending on the version, the relay connection cables, with the cable clips.
→
Tighten the cable glands making sure the cable in the housing is long enough to allow complete opening of
the housing.
→
Close the cover.
→
Fully screw the 4 screws.
→
Put the blanking strips on the housing.
22
Cables of the outputs
→
Before wiring the device insert the supplied cable clips into the slots of
the electronic board and of the 115/230 V AC power supply board if
the device has such a board.
English
E
→
Energize the device with a
115 V AC voltage.
Power supply cable
12-36 V DC or 115/230 V AC
Type 8025 UNIVERSAL
Installationandwiring