The following table shows the response times (10% to 90%) for the filters:
Filter
0
1
2
3
4
Filter 0: measured
flow rate
Figure 45:
Available filters
→
If you do not want to adjust another parameter, go to the "END" parameter of the Parameters menu and press
ENTER
to save the settings or not and go back to the Process level.
8.6.15
Resetting both totalizers
This parameter makes it possible to reset both totalizers.
Both totalizers are reset upon confirmation of "SAVE YES" when leaving the Parameters menu.
tOtAL
CUtOFF
Figure 46:
Diagram of the "TOTAL" parameter of the Parameters menu
The daily totalizer can be reset from the Process level. See chap. 8.5.
→
If you do not want to adjust another parameter, go to the "END" parameter of the Parameters menu and press
ENTER
to save the settings or not and go back to the Process level.
52
Response time
0,15 s
0,7 s
1,4 s
2,5 s
3,5 s
Filter 3
RESEt N
0......9
RESEt Y
→
Choose to reset the two totalizers or not.
→
Confirm
English
Filter
5
6
7
8
9
slow
Filter 6
fast
Type 8025 UNIVERSAL
Operatingandcommissioning
Response time
6 s
10 s
19 s
33 s
50 s
Filter 9