9.3.6
Resolution of problems linked to warning messages not
registered in the Information menu
Device
current
status leD
output
Ao1
any colour
4-20 mA depending
any colour
4-20 mA depending
64
output Do1
message
and/or Do2
displayed
and/or Do3
"PU H LIM" The message is displayed after
on
thresholds
"PU L LIM" The message is displayed after
on
thresholds
English
Type 8025 UNIVERSAL
Maintenanceandtroubleshooting
possible cause
the pulse value has been entered
(parameter "PU" of the transistor
output DO1) or upon validation
of the settings of the Parameters
menu.
The pulse value times the K factor
of the device is > 1000000.
The entered volume for a pulse is
too high.
the pulse value has been entered
(parameter "PU" of the transistor
output DO1) or upon validation
of the settings of the Parameters
menu.
The pulse value times the K factor
of the device is < 1.
The entered volume for a pulse is
too low.
recommended
action
→
Enter a lower
volume value for a
pulse. See chap.
8.6.8.
→
Enter a higher
volume value for a
pulse. See chap.
8.6.8.