To infuse a Clinician Activated Bolus during an infusion, perform the following steps:
During operation, press
1.
initiate the clinician bolus.
Enter relevant code and press
2.
3.
Enter the volume of the clinician bolus and press
to begin priming procedure. You may stop the
bolus at any time by pressing
4.
The volume for the bolus will be in ml, mg or µg
depending on the settings of the existing protocol
entered for the relevant infusion.
5.
The display screen will show a graph indicating the
Bolus is being infused. The graphical bar displays
the amount of the bolus to be infused (grey) and the
amount infused (green).
NOTE: once the bolus is completed, the pump will resume the original infusion
program.
NOTE: user can stop bolus at any time by pressing
NOTE: clinician bolus can also be set using 'Clinician Bolus' option on the menu.
It is recommended to select a protocol prior to starting a clinician bolus.

Automatic Bolus Rate Adjustment

The BodyGuard 595 Color Vision™ has a special feature which optimizes bolus rate
during infusion through narrow catheters such as Epidural catheters. The pump will
attempt to deliver the bolus at the pre-set bolus delivery rate. If the pump senses back
pressure above the pre-set occlusion pressure level during bolus delivery it will
BodyGuard 595 Color Vision™ Directions For Use
and then press
.
.
to stop the current infusion and
EPIDURAL
10:00
Clinician Bolus
Dose
5
mL
(Maximum
20mL)
Enter and Press
EPIDURAL
10:00
Clinician Bolus
Dose
5
mcg
(Maximum
40mcg)
Enter and Press
EPIDURAL
10:00
Bolusing
███████████████████
Bolus Dose
██████████████████████
20 of 400 mmHg
.
4 – Operation
-
70%
Key
-
70%
Key
-
70%
0.363 mL
5 mL
2 mL/hr
46