Operation; Explanation Of The Display - Atmos Thorax S 201 Operating Instructions Manual

Hide thumbs Also See for Thorax S 201:
Table of Contents

Advertisement

4.0 Operation

4.1 Explanation of the display

Flow
Display of the actual fistula
ATTENTION: This is the current value. In
the case of a discontinuous fi stula, the
value can decline temporarily to zero
although the fi stula still exists.
although the fi stula still exists.
Actual vacuum
Display of
the actual
vacuum value.
Target vacuum
Display of the
preadjusted
target vacuum
to which the
pump adjusts.
Start / Stop
therapy
Keylock activated
Keylock activated
System information
Keylock manual
Flow displayed as bubbles
Each additional coloured bubble represents an
additional fl ow.
None:
0 - < 30 ml/min
Green:
30 - < 60 ml/min
Yellow:
60 - < 630 ml/min
Orange:
630 ml - 1,00 l/min
Red:
> 1.00 l/min to maximum.
Up to 1,00 l/min the fl ow is displayed in ml/min.
Day/Night Mode
The ATMOS
S 201 Thorax has a day/night
®
mode, the device adjusts automatically to the
environmental lighting.
Under low ambient light conditions display has
dark background illumination.
Therapy progress
Changeover to the
graphic diagram
User settings
Setting button
Increases the
target vacuum
Setting button
Decreases the
target vacuum
19