Cpr Cable - Park House Healthcare RICHMOND ULTRA LOW BED Service Manual

Acute care bed
Table of Contents

Advertisement

CPR cable

Removal
1. Disconnect the mains cable from the electrical power supply
2. Ensure the back rest section is in the horizontal position
3. Snip the crimp off the end of the cable
4. Loosen and then remove the grub screw
5. Remove the CPR cable from the back rest actuator as described previously
Install
1. Thread the CPR cable through the CPR bracket
2. Place the grub screw onto the cable and tighten
3. Place a crimp on to the end of the cable and clamp until it is securly in place
4. Place the CPR cable into the back rest actuator and secure the lid by pressing down
firmly
5. Test the function and adjust the bolt on the handle where required
Richmond Ultra Low Bed – Service Manual
30