Castor - Park House Healthcare RICHMOND ULTRA LOW BED Service Manual

Acute care bed
Table of Contents

Advertisement

Castor

Removal
1. Disconnect the mains cable from the electrical power supply
2. Engage the brake system
3. Remove the head and foot boards from the bed
4. Place the bed on its side using a set of transporters
5. Remove the two bolts labelled ① at both ends of the brake bar
6. Remove the ② bolt from the castor
7. Adjust the bar labelled ③ so the castor can be freed
Install
1. Ensure the castor is in the locked position
2. Insert the castor through the beds bracket
3. Insert the bar through the castor ③
4. Secure the bar with the bolts labelled ①
5. Secure the castor with the bolt labelled ②
Richmond Ultra Low Bed – Service Manual
31