Troubleshooting - Park House Healthcare RICHMOND ULTRA LOW BED Service Manual

Acute care bed
Table of Contents

Advertisement

10. Troubleshooting

Symptom
Possible cause
No actuators will function
Nurse lock out in the locked position
Power has been disconnected and the battery is
flat
Patient handset not inserted
Damaged cable
Blown fuse
One actuator will not
Actuator has been disconnected
function
Damaged actuator cable
Assess handset for damage
Control box is clicking when handset is pressed
The bed will not function
Battery charge fault
from battery after charge
Brakes and/or castor(s)
Wear and tear / damaged
will not function correctly
Manual CPR will not
CPR cable too short
function
CPR cable cap missing from actuator
Richmond Ultra Low Bed – Service Manual
Action
Unlock the nurse lockout
Connect the bed to the power supply and allow to charge for
24hrs
Insert the handset into the nurse handset
Replace the damaged cable
Replace the fuse
Reconnect the actuator and secure the security clip
Replace the actuator
Replace handset
Actuator exceeded its limit switch, replace actuator
Replace battery
Replace the damaged castor
Adjust the cable length accordingly
Replace the CPR actuator cap
33