Part Replacement; Mains Lead - Park House Healthcare RICHMOND ULTRA LOW BED Service Manual

Acute care bed
Table of Contents

Advertisement

9. Part replacement

CAUTION! Ensure the mains supply is isolated prior to replacing any parts

Mains lead

Removal
1. Disconnect the mains cable from the electrical power supply
2. Remove the trunking lid on the lower platform section
3. Remove the 2 screws connecting the control box to the bed frame
4. Lift the control box
5. Remove the mains cable from trunking
6. Using a terminal screw driver, press the red clip down
7. Pull the mains lead away from the control box
Install
1. Insert the mains lead into the control box, the cable can only be inserted one way
2. Ensure the securing clip on the mains lead engages with control box
3. Pull the mains lead to ensure the cable is secure
4. Secure the control box back to the bed frame
5. Place the mains cable in the trunking and then secure the top back on the trunking
6. Secure the cable along the lower platform with the grey C clips provided
Richmond Ultra Low Bed – Service Manual
20