6.4.1 Fixed Or Portable Operation Selection
The first menu is for setting the MeshGuard for fixed or portable operation. In Portable Mode, STEL, TWA, and
Peak readings are included.
"SEt" and " Ptb" flash in alternation, to indicate that the MeshGuard can now be set as fixed or portable. The default
value is "fixed."
Press [Y/+] to enter and change the setting, [MODE] to exit and return to Detection Mode, or [N/-] to go to the next
menu.
The LCD displays the current value (enabled or disabled). Change the value if necessary by pressing [Y/+] or [N/-].
Press [MODE} to enter your selection. Press [Y/+] to accept at this screen.
Note: "dIS" means fixed (Portable Mode disabled) and "En" means portable (Portable Mode enabled).
6.4.2 Radio Enable/Disable
The next menu is for setting turning the MeshGuard's radio on or off.
"SEt" and " rdo" flash in alternation, to indicate that the MeshGuard radio can be turned on (enabled) or off
(disabled). The default value is "on."
Press [Y/+] to enter and change the setting, [MODE] to exit and return to Detection Mode, or [N/-] to go to the next
menu.
The LCD displays the current value (enabled or disabled). Change the value if necessary by pressing [Y/+] or [N/-].
Press [MODE} to enter your selection. Press [Y/+] to accept at this screen.
Note: "dIS" means disabled (radio off) and "En" means on (radio on).
6.4.3 Std Enable or Rtr Enable
Press [Y/+] to toggle the setting from Std to Rtr, and vice versa. Refer to "Operation Modes" on page 13.
MeshGuard User's Guide
32