Recommended Separation Distances - Planmeca Emerald User Manual

Hide thumbs Also See for Emerald:
Table of Contents

Advertisement

Immunity Test IEC 60601 Test
Level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to
80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2.5 GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio,
AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment
due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the
Emerald is used exceeds the applicable RF compliance level above, the Emerald should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating the Emerald.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

Recommended Separation Distances

Recommended separation distances between portable and mobile RF communications equipment and the
Emerald
The Emerald is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the Emerald can help prevent electromagnetic interference by maintaining
a minimum distance between portable and mobile RF communications equipment (transmitters) and the Emerald
as recommended below, according to the maximum output power of the communications equipment.
Rated maximum
Separation distance according to frequency of transmitter m
output power of
transmitter W
0.01
0.1
1
10
100
15698200.B
Compliance
Level
Not
applicable
3 V/m
150 kHz to 80 MHz
0.12
0.38
1.2
3.8
12
Electromagnetic Environment - guidance
Portable and mobile RF communications equipment should
be used no closer to any part of the Emerald, including cables,
than the recommended separation distance calculated from
the equation applicable to the frequency of the transmitter.
Recommended separation distance
where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer and
d is the recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined
by an electromagnetic site survey,
compliance level in each frequency range.
occur in the vicinity of equipment marked with the following
symbol:
80 MHz to 800 MHz
0.12
0.38
1.2
3.8
12
Planmeca Emerald Specifications
should be less than the
a
Interference may
b
800 MHz to 2.5 GHz
0.23
0.73
2.3
7.3
23
51