Setup Screen

Use the Setup screen to set the restoration type, occlusal data type, material, and tooth library. If you open an
existing restoration, many of these settings may already be selected.
For diagnostic cases, orthodontic aligners, or when sending a large restorative case (without drawing margins) to a
laboratory, select Full Arch as the restoration type and click Scan to proceed.
The Setup screen must be filled out for restorative cases.
The Setup screen requires that the settings be made in a particular order because certain parameters depend upon
previous selections. Always define a restoration in the following order:
1
Select a restoration site by clicking on the anatomical model. If no teeth have been
selected, the anatomical model is closed until the mouse cursor is moved near it.
The currently selected tooth is orange. If any other teeth are part of this restoration file,
they are green.
For bridges, select the abutments and the pontic(s).
You can do up to 16 restorations on the same arch at one time.
To change the tooth numbers, see the Romexis User Manual.
2
Choose the restoration type.
3
Select the opposing scan type.
The majority of exported case recipients will require Buccal/Opposing cases.
Bite registration cases can only be sent to another PlanCAD system.
For Scan Only cases, the library and material selections can be randomly selected.
4
Specify a tooth library. See the PlanCAD manual for more information.
5
Specify a material. See the PlanCAD manual for more information.
6
Specify a translucency (not applicable for all materials).
7
Specify a shade. See the PlanCAD manual for more information.
Repeat these steps if you have multiple restorations.
8
Click the Scan screen or click Next.