4700 Refrigerated Sampler

3.5.9 Run Continuously?

Multiple Bottles Only – Sample programs can
run indefinitely by selecting YES at the RUN
CONTINUOUSLY? screen. Continuous
sampling resets the distribution when the
distribution sequence is complete. That is,
when the last bottle/set is reached, the next
bottle/set is the first bottle/set.
In this mode, the sampler assumes that any
filled bottles are replaced and ready to receive
samples while the program is running, thus the
pacing interval is never interrupted.
If you select NO, the sampler will run until it
completes the distribution sequence, at which
time it stops the program and reports
PROGRAM DONE. The sampler will wait in
this state until the bottles have been emptied
and the program is restarted.
If your sampler is configured for composite
sampling, refer to section 3.5.16 for instructions
on continuous sampling.
3.5.10 Enter Sample Volume Use Defined
Sample
At this screen you have the option to enter a
sample volume (factory default is 200 ml) or use
a manually defined sample volume based on
pump counts.
If you choose ENTER SAMPLE VOLUME, the
programming displays will step through a
series of screens that let you enter the sample
volume. When running a program, the sampler
3-34