Select the SUCTION LINE option and
press Enter.
2.
Press the Number buttons to enter the
suction line length.
3.
The sampler creates pump tables before
returning to the SELECT OPTION <– –>
screen.
When accurate, repeatable sample volumes are
important, use care to cut the suction line and
enter the correct measurement.
If using English units of measure, cut the
suction line to the nearest whole foot. When
using metric units of measure, cut the suction
line to the nearest 0.1 meter. The sampler
displays units of measure appropriate for the
selected language (see section 3.3.11).

3.3.4 Liquid Detector

Select this option to turn liquid detection on or
off. This setting also allows you to configure
suction line rinses and sample retries.
1.
Select the LIQUID DETECTOR option
and press Enter.
2.
SUCTION LINE LENGTH:
___ ft
(3-99)
PLEASE WAIT!
CREATING PUMP TABLES
SELECT OPTION: (<-->)
LIQUID DETECTOR
USE LIQUID DETECTOR?
YES
Section 3 Programming
NO
3-7