Teledyne 4700 Installation And Operation Manual

Teledyne 4700 Installation And Operation Manual

Refrigerated samplers
Hide thumbs Also See for 4700:
Table of Contents

Advertisement

Quick Links

Teledyne Isco,
Inc.
Project Name:
Clinton Wastewater Treatment Systems
Improvements, Contract
C
Project Owner: City of Clinton, Iowa
Engineer:
Contractor:
HDR
Attn.: Chip Haas
8404
Indian Hills Dr.
Omaha, NE
68114
Ph. 402-399-1333
Gridor Construction, Inc.
Attn.: Gregg Schreiner
3990
2-rh
St. SE
Buffalo, MN
55313
Ph. 763-746-9075
I
\~
Purchase Order:
2010-03/11741
Manufacturer:
Teledyne Isco, Inc.
Attn.: Bill Soulliere
4700
Superior St.
Lincoln, NE
68504
Ph. 1-866-298-6174
Teledyne Isco
Sales Order No.:
6147853
Manufacturer's
Representative: Harry Ferris Company
Attn.: Jim Ferris
7404
Hampshire Dr. NE
Cedar Rapids, IA
52402
Ph. 319-377-1226

Advertisement

Table of Contents
loading

Summary of Contents for Teledyne 4700

  • Page 3 4700 Refrigerated Sampler Installation and Operation Guide Part #69-4703-086 of Assembly #60-4704-051 Copyright © 2006. All rights reserved, Teledyne Isco, Inc. Revision C: May 24, 2007...
  • Page 8 Although Isco designs reliability into all equipment, there is always the possibility of a malfunction. This manual may help in diagnosing and repairing the malfunction. If the problem persists, call or E-mail the Teledyne Isco Customer Service Department for assistance. Contact information is provided below. Simple difficulties can often be diagnosed over the phone.
  • Page 9 Isco 4700 Refrigerated Sampler Isco 4700 Refrigerated Sampler Safety Information The Isco 4700 Refrigerated Sampler is a “definite purpose” device, intended for use only with compatible Isco equipment. Except as described in this manual, do not use this product with any other manufacturers’...
  • Page 10 Isco 4700 Refrigerated Sampler General Warnings Before installing, operating, or maintaining this equipment, it is imperative that all hazards and preventive measures are fully understood. While specific hazards may vary according to location and application, take heed in the following general...
  • Page 11 Isco 4700 Refrigerated Sampler The equipment and this manual use symbols used to warn of Hazard Symbols hazards. The symbols are explained below. Hazard Symbols Warnings and Cautions The exclamation point within the triangle is a warning sign alerting you of important instructions in the instrument’s technical reference manual.
  • Page 15: Section 1 Introduction

    • Features, page 1-2 • Specifications, page 1-4 1.1 Sampler Overview The 4700 Refrigerated Sampler is an automated wastewater sampler that can collect liquid samples from a variety of sources including open channels, sewers, and storm pipe conduits. This indoor/outdoor use sampler is made of durable UV-resistant plastic that provides exceptional resistance to corrosion and weathering.
  • Page 16: Features

    Isco 4700 Refrigerated Sampler Section 1 Introduction 1.2 Features The 4700 Series sampler features are identified in Figure 1-1 and described in Table 1-1. Front View Left View Back View Figure 1-1 4700 Sampler features...
  • Page 17 Isco 4700 Refrigerated Sampler Section 1 Introduction Table 1-1 4700 Sampler Features Item Name Description (Fig. 1-1) Control Panel Cover Protects the control panel display and keypad. Control Panel Supports user control and programming of sampler operation. Discharge Tube Carries liquid to the sample bottle. For 24, 4, 2, and 10 liter one-bottle configurations, the discharge tube is routed through the distributor arm.
  • Page 18: Specifications

    Isco 4700 Refrigerated Sampler Section 1 Introduction 1.3 Specifications Table 1-2 Specifications 130 × 72 × 84 cm 51 × 28 × 33 in Size (H×W×D): Weight (empty): 72 kg 159 lb Bottle Configurations: 12 different configurations available: • 24 bottles, 1-liter PP or 350 ml glass;...
  • Page 19 Isco 4700 Refrigerated Sampler Section 1 Introduction Table 1-2 Specifications (Continued) Controller Enclosure Rating: IP67 NEMA 4X, 6 Program Memory: Non-volatile ROM (Flash) Flow Meter Signal Inputs: 5 to 15 volt DC pulse or 25 millisecond isolated contact closure for Isco flow meters.
  • Page 20: Section 2 Installation

    If there is any Sampler damage to the shipping carton or any components, contact the shipping agent and Teledyne Isco (or its authorized represen- tative) immediately. WARNING If there is any evidence that the sampler has been damaged in shipping, do not plug it into an AC power source.
  • Page 21 Isco 4700 Refrigerated Sampler Section 2 Installation WARNING If this product is used in a manner not specified in this manual, the protection provided by the equipment may be impaired. WARNING This product has not been approved for use in hazardous locations as defined by the National Electrical Code.
  • Page 22: Minimum Space Requirements

    Isco 4700 Refrigerated Sampler Section 2 Installation • Liquid sample collection point – It is best to keep the distance between the sampler pump and the collection point as short as possible. The shorter distance will increase pumping efficiency and pump tubing life. Also, the pump inlet should be located above the liquid surface.
  • Page 23: Installing A Bottle Kit

    2.4 Installing a Bottle Kit The 4700 sampler can hold 1, 2, 4, and 24 bottles. Each of these bottle configurations are supplied as a kit (see Table 2-1), which is normally ordered with the sampler or when you desire to change the bottle configuration.
  • Page 24: Installing The Distributor Arm And Discharge Tube

    Isco 4700 Refrigerated Sampler Section 2 Installation If you have selected a multi-bottle kit, first install the distributor arm and discharge tube (Section 2.4.1), then refer to the instruc- tions for your selected bottle configuration: • 24 Bottle configurations – section 2.4.2 •...
  • Page 25 Then, retest the tube at other arm positions. Note You can order replacement discharge tubes from Teledyne Isco. Order part number 60-9003-279 for a single tube. You can also order bulk tubing rolls from which you can cut to 20 inches (51 cm).
  • Page 26: Insert The Keyed End First So That It Faces The Back Of The Refrigerated Compartment

    Isco 4700 Refrigerated Sampler Section 2 Installation Note There are four metal guides along the sidewall of the refrigera- tor, two on each side. The front two guides fit along the left and right sides of the rack. The back two guides run along the top of the rack to prevent the rack from tipping forward.
  • Page 27: Bottles

    Isco 4700 Refrigerated Sampler Section 2 Installation 2.4.3 4 Bottles The four bottle configuration uses a locating base to hold and align the bottles. To install the base and bottles: 1. Slide the locating base into the bottom of the refrigerated compartment (Figure 2-5).
  • Page 28: Bottle

    Isco 4700 Refrigerated Sampler Section 2 Installation 2.4.5 1 Bottle If your one-bottle configuration has a 10 liter (2.5 gallon) polyeth- ylene, glass, or ProPak bottle, use a locating base to hold and align the bottle. If your one-bottle configuration has a 20 liter (5 gallon) glass or polyethylene bottle, simply place the bottle in the center of the refrigerator;...
  • Page 29: Attaching The Suction Line

    The sampler uses a -inch ID suction line of Line lengths from 3 to 99 feet. Teledyne Isco offers vinyl or FEP-lined suction lines. The FEP-lined tubing has a polyethylene jacket to protect it from kinks and abrasions.
  • Page 30: Vinyl Suction Line

    Isco 4700 Refrigerated Sampler Section 2 Installation When accurate, repeatable sample volumes are important, use care to cut the suction line to the correct increments: • When the sampler is using metric units of measure, cut the line in 0.1 m increments. For instance, cut 4.2 m, not 4.26 m.
  • Page 31: Fep-Lined Suction Line

    Isco 4700 Refrigerated Sampler Section 2 Installation Figure 2-8 Releasing a tube coupler clamp 2.5.2 FEP-lined Suction Line Inserting a tube coupler into FEP-lined suction line will damage the thin lining. Instead, refer to Figure 2-8 and the instructions below to attach FEP-lined suction line to the pump tube: 1.
  • Page 32: Attaching A Strainer

    Isco 4700 Refrigerated Sampler Section 2 Installation 2.6 Attaching a Strainer To select the right strainer for your application, see Table 2-2 and Figure 2-10. To attach the strainer to the suction line, carefully screw the strainer’s threaded connector into the suction line. If attaching the strainer to an FEP-lined suction line, heat the end of the suction line before threading the strainer into the line.
  • Page 33: Alternative To Strainers

    Isco 4700 Refrigerated Sampler Section 2 Installation 2.6.1 Alternative to Strainers When sampling from high velocity streams with heavy sus- pended solids, some field investigations suggest that more repre- sentative samples are obtained without the strainer. Consider attaching a short piece of thin walled aluminum tubing to the end of the suction line;...
  • Page 34: Standard Isco Instrument Connections

    • Power the flow measurement device. To connect either of these instruments, use the optional 7.6 m (25 ft) connect cable, 69-4704-043 (Figure 2-12). To sampler To Isco instrument Figure 2-12 4700 Sampler to Isco flow meter cable 2-15...
  • Page 35: Other Device Connections

    All of these types of connections use the optional unterminated connect cable, P/N 68-4700-020 (Figure 2-13). This 3 m (10 ft) cable is unterminated at the device end to allow wired connec- tions to compatible circuits. Refer to Table 2-3 for wire identifi- cation and connection details.
  • Page 36: Unterminated Connect Cable

    Isco 4700 Refrigerated Sampler Section 2 Installation Table 2-3 Unterminated Connect Cable Sampler Pin Wire Color Signal Name Input/Output Parameters/Comments Black 12 VDC Power Output +12.5 VDC, 1 ampere maximum. Power for external devices. White Ground Common ground Green Flow Pulse...
  • Page 37: Connecting Power

    Isco 4700 Refrigerated Sampler Section 2 Installation 2.9 Connecting Power WARNING Before connecting the sampler to an AC power source, be familiar with the Electrical Requirements listed on page v at the front of this manual. The factory assembles the sampler for either 100VAC/50 Hz, 115 VAC/60 Hz, or 230 VAC/50 Hz.
  • Page 38: Section 3 Programming

    Isco 4700 Refrigerated Sampler Section 3 Programming In this section… • Control Panel Description • Getting Started, page 3-2 • Configuring the Sampler, page 3-3 • Sampling Program Overview, page 3-10 • Programming Steps, page 3-13 • Programming Examples, page 3-20 3.1 Control Panel...
  • Page 39: Getting Started

    Isco 4700 Refrigerated Sampler Section 3 Programming Table 3-1 Control Buttons (Continued) Icon Name Description Previous Selects the previous menu option left or above the current choice. Next Selects the next menu option right or below the cur- rent choice.
  • Page 40: Configuring The Sampler

    Isco 4700 Refrigerated Sampler Section 3 Programming 3.3 Configuring the Before operating the 4700 Sampler, configure the sampler software. Doing so will set the time and date, and allow the Sampler sampler controller to correctly use the hardware and external connections.
  • Page 41: Bottle And Sizes

    Isco 4700 Refrigerated Sampler Section 3 Programming 3.3.2 Bottle and Sizes Select this option to specify the installed bottle configuration (section 2.4). The sampler uses this setting to determine available bottle options and control the operation of the dis- tributor.
  • Page 42: Liquid Detector

    Isco 4700 Refrigerated Sampler Section 3 Programming 3.3.4 Liquid Detector Select this option to turn liquid detection on or off. This setting also allows you to configure suction line rinses and sample retries. 1. Select the LIQUID DETECTOR option and press Enter.
  • Page 43: Refrigeration

    Isco 4700 Refrigerated Sampler Section 3 Programming 2. Select YES for unlimited enable/disables. Select NO to REPEATABLE ENABLE? allow only one enable, after which the sampler will remain enabled until the end of the program sequence. 3. If your application requires a sample when the sampler SAMPLE AT ENABLE? becomes enabled, select YES.
  • Page 44: Tubing Life

    Isco 4700 Refrigerated Sampler Section 3 Programming • PROGRAM COMPLETED – This event occurs when a running sampling program has finished. This alarm output stays on as long as the run time display reads PROGRAM COMPLETED. • PUMP FAILURE – This event occurs when the sampler has detected a pump failure.
  • Page 45: Program Lock

    Check the condition of the pump tube regularly and if the tube shows signs of fatigue or wear, replace it immediately. A properly maintained sampler will pro- vide years of reliable service that is expected of a Teledyne Isco Sampler. 3.3.10 Program Lock Select this option to turn the Program Lock feature on or off, and to change the password.
  • Page 46: Language

    ____ them again. Should you forget the password, contact Teledyne Isco’s Cus- tomer Service department for assistance. 3.3.11 Language Select this option to change the display language and display units for length measurements.
  • Page 47: Sampling Program Overview

    Isco 4700 Refrigerated Sampler Section 3 Programming 3.4 Sampling Program The sampling program controls how often sample events occur and what should take place at each event. A sample event may Overview happen when: • you start a flow paced program that is programmed to Sample at Start (section 3.3.5).
  • Page 48: Time Paced Sampling Program Flow Chart

    Isco 4700 Refrigerated Sampler Section 3 Programming PROGRAM CONFIGURE VIEW LOG TIME PACED 3.5.1 FLOW PACED SAMPLE EVERY 3.5.2 __ HOURS, __ MINUTES One Bottle Multiple Bottles __ BOTTLES PER __ SAMPLES PER 3.5.4 3.5.8 SAMPLE EVENT (1-max) BOTTLE (1-max) RUN CONTINUOUSLY? 3.5.9...
  • Page 49: Flow Paced Sampling Program Flow Chart

    Isco 4700 Refrigerated Sampler Section 3 Programming PROGRAM CONFIGURE VIEW LOG TIME PACED 3.5.1 FLOW PACED SAMPLE EVERY 3.5.3 __ PULSES (1-9999) One Bottle Multiple Bottles __ BOTTLES PER 3.5.4 SWITCH BOTTLES EVERY SAMPLE EVENT (1-max) 3.5.6 __ HOURS, __ MINUTES SWITCH ON TIME 3.5.5...
  • Page 50: Programming Steps

    Isco 4700 Refrigerated Sampler Section 3 Programming 3.5 Programming Steps To begin programming from the Main Menu screen, use the Next or Previous button to select PROGRAM. Press the Enter button to display the first programming screen. Refer to Figures 3-3 or 3-4 and the following descriptions.
  • Page 51: Flow Pulse Intervals At Various Input Currents

    Isco 4700 Refrigerated Sampler Section 3 Programming If the flow measurement device sends flow rate data via a 4-20 mA current loop instead of flow pulses, the sampler converts this analog current to flow pulses representative of a volume. The sampler assumes that the current is linear from 4 mA at zero flow to 20 mA at the full-scale flow rate.
  • Page 52 Isco 4700 Refrigerated Sampler Section 3 Programming Time Unit Factor time seconds (cfs, m s, etc.) 0.08333 minutes (gpm, l/m, etc.) hours (gph, m h, etc.) days (mgd, m d, etc.) 7200 3. Divide the sample interval (I ) by the result of step 2.
  • Page 53: Bottles Per Sample Event

    Isco 4700 Refrigerated Sampler Section 3 Programming 3.5.4 __ Bottles per Sample Multiple Bottles Only – The sampler places a sample volume in Event one bottle or sets of bottles at each sample event. Use this screen to enter the number of bottles to repeat the sample volume at each sample event.
  • Page 54: Run Continuously

    Isco 4700 Refrigerated Sampler Section 3 Programming Distribution scheme with three samples per bottle. This Samples per Bottle feature can be combined with the Bottles per Sample Event (section 3.5.4) to build more complex distribution schemes, sometimes known as multiple bottle com- positing.
  • Page 55: Volumes Dependent On Flow

    Isco 4700 Refrigerated Sampler Section 3 Programming Note When using defined samples, be sure to DEFINE SAMPLE using the Calibrate button. See the example “Defining the Sample Volume” on page 3-21. Note Selecting USE DEFINED SAMPLE will disable bottle overflow detection.
  • Page 56: Suction Head

    Figure 3-5 Suction Head Height Applications that collect samples from a pressurized line are an exception. Teledyne Isco offers a pressurized line option for the sampler. Contact the factory for more information. When using Teledyne Isco’s pressurized line option, disable liquid detection and enter “1”...
  • Page 57: First Sample At

    Isco 4700 Refrigerated Sampler Section 3 Programming Note If you enter a “First Sample At” or Start Flow Count At” month less than the current month, the sampler interprets the start time as next year’s date. 3.5.19 First Sample At: Time Paced Only –...
  • Page 58 Isco 4700 Refrigerated Sampler Section 3 Programming Example: Defining the Sample Volume If you have programmed the sampler to USE DEFINED deposit it into the bottle. To measure the delivered SAMPLE (section 3.5.10), follow the steps in this volume, you can replace the bottle with a example to define the volume using pump counts.
  • Page 59 Isco 4700 Refrigerated Sampler Section 3 Programming Example: Time Paced Sampling Program This example shows how to program the sampler to 7. The sampler displays the option to enter a sample collect a 500 ml sample every hour for twenty-four volume or use the defined volume.
  • Page 60 Isco 4700 Refrigerated Sampler Section 3 Programming Example: Flow Paced Sampling Program This example shows how to program the sampler to 5. The sampler displays the Bottle Switch screen. Our collect a 100 ml sample every 10,000 gallons of liquid program requires a bottle switch every 24 hours.
  • Page 61 Isco 4700 Refrigerated Sampler Section 3 Programming Flow Paced Sampling Program, continued 12. The sampler displays the Start Flow Count screen. 14. The sampler displays the Programming Sequence Type “00” and press Enter, then type “00” and Complete screen for four seconds, then advances...
  • Page 62 Isco 4700 Refrigerated Sampler Section 3 Programming Example: Flow-proportional Constant Time Variable Volume Program This example shows how to program the sampler to 5. The sampler displays the Volume Dependent On collect flow-proportional samples volumes at fixed time Flow screen. Select YES and press Enter.
  • Page 63: Section 4 Operation

    Isco 4700 Refrigerated Sampler Section 4 Operation In this section… • Start a Sampler Program • Pause or Stop a Running Program, page 4-3 • Post Sampling Activities, page 4-3 • View the Log, page 4-3 • Retrieve the Temperature Log, page 4-5 •...
  • Page 64: The Run State

    Isco 4700 Refrigerated Sampler Section 4 Operation If the program is set to use a start time, the sampler will wait for the specified time and date. During this delay, the display for time paced programs will show: FIRST SAMPLE AT:...
  • Page 65: Pause Or Stop A Running Program

    Isco 4700 Refrigerated Sampler Section 4 Operation 4.2 Pause or Stop a Press the red Stop button to pause a running program. The sampler will display: Running Program RESUME IN MM:SS VIEW LOG HALT The top line of the display shows a five minute count down timer.
  • Page 66 Isco 4700 Refrigerated Sampler Section 4 Operation The log generally contains the following information: 1. The log reports the number of successful samples out of the COLLECTED total number programmed. ___ OF ___ SAMPLES 2. If samples were not collected, the log reports the number of MISSED SAMPLES: ___ missed samples and the error message.
  • Page 67: Retrieve The Temperature Log

    Isco 4700 Refrigerated Sampler Section 4 Operation Table 4-1 4700 Errors Message Description The sampler was unable to detect liquid. NO LIQUID DETECTED After the sampler detected liquid and while the NO MORE LIQUID sample was being taken, the liquid detector stopped detecting liquid.
  • Page 68: Sample Temperature Report

    Isco 4700 Refrigerated Sampler Section 4 Operation To sampler To computer Figure 4-1 4700 Sampler to RS-232 serial communication cable 4700 Sampler,Model,4700,ID,1158325658, Current Time,38980.62200, Run Time,28126.00000, First Sample Time,38978.55000, 38980.61250, 1.8*, 38980.61319, 2.1, 38980.61389, 2.4, 38980.61458, 2.6, 38980.61528, 2.9, 38980.61597, 3.0,...
  • Page 69: Remove Sample Bottles

    Isco 4700 Refrigerated Sampler Section 4 Operation 4.3.3 Remove Sample After the program is complete, the bottles can be removed and Bottles prepared for the laboratory. Gain access to the bottles by releasing the door latch and swinging the door open. The bottles can then be removed from the rack or locating base.
  • Page 70: Grab Samples

    Isco 4700 Refrigerated Sampler Section 4 Operation 4.4 Grab Samples Grab samples let you take a single sample on demand and col- lecting it in an external container. You can collect a grab sample while the sampler is running a program, paused, or at the Main Menu screen.
  • Page 71: Operate The Pump Manually

    (section 3.3.3). These steps also require a graduated cylinder to measure the delivered sample volume. Note Teledyne Isco offers a 1000 ml graduated cylinder. Order part number 299-0020-00. To calibrate the sample volume: 1. At the Main Menu, press the Calibrate button.
  • Page 72: Section 5 Maintenance

    This section describes how to maintain and troubleshoot your sampler. If you think your sampler requires repair, or if you have questions concerning its operation or maintenance, contact your authorized Isco service facility or Teledyne Isco’s Technical Service Department: Phone: (800) 228-4373 (USA, Canada, Mexico)
  • Page 73 Isco 4700 Refrigerated Sampler Section 5 Maintenance WARNING The sampler has a safety interlock that prevents the pump from operating when the pump housing band is open. DO NOT tamper with the pump housing and band. The pump is extremely powerful. The pump can injure you severely if the sampler activates the pump while you are working on it.
  • Page 74: Pump Tube Replacement

    Isco 4700 Refrigerated Sampler Section 5 Maintenance Figure 5-2 Rollers (A) and guides (B) on the pump rotor b. Pump Rotor – Look for debris build-up on the pump roller or guide surfaces (Figure 5-2). Clean when needed (section 5.1.3).
  • Page 75 Isco 4700 Refrigerated Sampler Section 5 Maintenance 2. Loosen the Liquid Detector Cover (G) by unscrewing the large black knob (I). 3. Unlatch the Pump Housing Band (K). 4. Pull the tube (A) away from the bulkhead fitting (B) and disconnect the tubing connector and suction line.
  • Page 76: Cleaning The Pump Rollers

    Isco 4700 Refrigerated Sampler Section 5 Maintenance Example: Resetting the Pump Tube Alarm 1. From the Main Menu screen select CONFIGURE. 5. You can modify the pump tube alarm setting to tailor it to your needs. The factory default setting is...
  • Page 77: Sampler Cleaning Guidelines

    Isco 4700 Refrigerated Sampler Section 5 Maintenance Figure 5-3 Wetted Parts Strainer (316 stainless steel, Bulkhead Fitting (316 stain- polypropylene, or CPVC) less steel) Suction Line (vinyl) Discharge Tube (silicone) Tubing Connector (316 stain- Bottle (glass, polypropylene, less steel) or polyethylene) Pump Tube (silicone) 5.1.6 Sampler Cleaning...
  • Page 78: Test 'Ram

    3. When finished, the sampler displays the test results. 'RAM' PASSED If the sampler reports ‘RAM’ FAILED contact your authorized Isco service facility or the Teledyne Isco factory. 5.2.2 Test ‘ROM’ The TEST ‘ROM’ diagnostic tests the sampler’s Read-Only Memory which holds the sampler software. It calculates a checksum total of the data held in this memory and compares it with a known value.
  • Page 79: Test Display

    · garbled text · poor contrast. If any of these or other faulty display conditions exist, contact your authorized Isco service facility or the Teledyne Isco factory. 5.2.4 Test Keypad The TEST KEYPAD diagnostic tests the sampler’s keypad. While the test is active, the display reports the button name of any button you press.
  • Page 80: Test Distributor

    TEST PUMP diagnostic again. If you don’t find obstructions to pump rotor movement, contact your authorized Isco service facility or the Teledyne Isco factory. 5.2.6 Test Distributor The TEST DISTRIBUTOR diagnostic tests the sampler’s dis-...
  • Page 81: Test Flow Meter Port

    Isco service facility or the Teledyne Isco factory. They will be able to assist you with obtaining a test plug or making the necessary jumper connec- tions to run the test.
  • Page 82: Input Voltage

    PCB TEMPERATURE: __._ C If you suspect that the internal heater is not working, contact your authorized Isco service facility or the Teledyne Isco factory. 5.2.10 Input Voltage The INPUT VOLTAGE diagnostic displays the Main CPU board’s DC voltage supplied by the power module. When this test is started, the sampler will briefly display the input power voltage.
  • Page 83 Isco 4700 Refrigerated Sampler Section 5 Maintenance Note Repeated use of the Loaded Battery Volts diagnostic will shorten the battery life. To display the loaded battery voltage: 1. Select the LOADED BATTERY VOLTS option and press SELECT DIAG: (<-->) Enter.
  • Page 84: Temperature Verification

    Isco 4700 Refrigerated Sampler Section 5 Maintenance 5.3 Temperature Should there be a need to verify the temperature readings dis- played by the control panel, it is important to read the temper- Verification ature as close as possible to the sensor inside the refrigerator.
  • Page 96: Appendix A Replacement Parts

    A.1 Overview Replacement parts are called out in the following illustrations. Refer to the tables to determine the part number for the item. Replacement parts can be purchased by contacting Teledyne Isco’s Customer Service Department. Teledyne Isco, Inc. Customer Service Department P.O.
  • Page 97 Isco 4700 Refrigerated Sampler Appendix A Replacement Parts...
  • Page 98 Isco 4700 Refrigerated Sampler Appendix A Replacement Parts...
  • Page 99 Isco 4700 Refrigerated Sampler Appendix A Replacement Parts...
  • Page 100 Isco 4700 Refrigerated Sampler Appendix A Replacement Parts...
  • Page 101 Isco 4700 Refrigerated Sampler Appendix A Replacement Parts...
  • Page 102 Isco 4700 Refrigerated Sampler Appendix A Replacement Parts...
  • Page 103 Isco 4700 Refrigerated Sampler Appendix A Replacement Parts...
  • Page 104 Isco 4700 Refrigerated Sampler Appendix A Replacement Parts...
  • Page 105 Isco 4700 Refrigerated Sampler Appendix A Replacement Parts A-10...
  • Page 106 Isco 4700 Refrigerated Sampler Appendix A Replacement Parts A-11...
  • Page 107 Isco 4700 Refrigerated Sampler Appendix A Replacement Parts A-12...
  • Page 108 Isco 4700 Refrigerated Sampler Appendix A Replacement Parts A-13...
  • Page 109 Isco 4700 Refrigerated Sampler Appendix A Replacement Parts A-14...
  • Page 110 Isco 4700 Refrigerated Sampler Appendix A Replacement Parts A-15...
  • Page 111 Isco 4700 Refrigerated Sampler Appendix A Replacement Parts A-16...
  • Page 112 Isco 4700 Refrigerated Sampler Appendix A Replacement Parts A-17...
  • Page 113 Isco 4700 Refrigerated Sampler Appendix A Replacement Parts A-18...
  • Page 114 Isco 4700 Refrigerated Sampler Appendix A Replacement Parts A-19...
  • Page 115 Isco 4700 Refrigerated Sampler Appendix A Replacement Parts A-20...
  • Page 116 Isco 4700 Refrigerated Sampler Appendix A Replacement Parts A-21...
  • Page 117 Isco 4700 Refrigerated Sampler Appendix A Replacement Parts A-22...
  • Page 118 Isco 4700 Refrigerated Sampler Appendix A Replacement Parts A-23...
  • Page 119 Isco 4700 Refrigerated Sampler Appendix A Replacement Parts A-24...
  • Page 120 Isco 4700 Refrigerated Sampler Appendix A Replacement Parts A-25...
  • Page 121 Isco 4700 Refrigerated Sampler Appendix A Replacement Parts A-26...
  • Page 122 Isco 4700 Refrigerated Sampler Appendix A Replacement Parts A-27...
  • Page 123: Appendix B General Safety Procedures

    Isco 4700 Refrigerated Sampler Appendix B General Safety Procedures The safety of the personnel who use the sampler is a critical con- sideration. The following procedures, applicable to working in and around manholes and sewers, are those used by Black &...
  • Page 124: Planning

    Isco 4700 Refrigerated Sampler Appendix B General Safety Procedures This applies to hand tools as well as stones, gravel and other objects. • Sharp Edges–Sharp edges of items in or near a manhole may cause cuts or bruises. • Lifting Injuries–Unless proper tools are used to remove manhole covers, back injuries or injuries to hands or feet may result.
  • Page 125: Traffic Protection

    Isco 4700 Refrigerated Sampler Appendix B General Safety Procedures A person should never enter a manhole unless he is wearing per- sonal safety equipment, including a safety harness and a hard hat. Two persons should be stationed at the surface continuously while anyone is working inside a manhole, to lift him out if he is overcome or injured.
  • Page 126: Emergencies

    Isco 4700 Refrigerated Sampler Appendix B General Safety Procedures B.4.4 Emergencies Every member of the crew should be instructed on procedures to be followed in cases of an emergency. It is the duty of each crew chief to have a list of emergency phone numbers, including the nearest hospital and ambulance service, police precinct, fire station, and rescue or general emergency number.
  • Page 127 Isco 4700 Refrigerated Sampler Appendix B General Safety Procedures Whitney, oxygen is equal to 12.2%t. Many hikers go there, and to higher peaks as well. After adequate acclimation, they may climb to the top of Mt. Everest, where oxygen is equal to only 6.7%.
  • Page 128: Hazardous Gases

    Isco 4700 Refrigerated Sampler Appendix B General Safety Procedures tions. A supposed safety device that is unreliable is worse than none at all. “Remember that the nose fails, too, when it comes to sens- ing dangerous concentrations of H “Various other toxic gases have been mentioned in some publications.
  • Page 129 Isco 4700 Refrigerated Sampler Appendix B General Safety Procedures Table B-1 Hazardous Gases (Continued) Chemical Common Specific Physiological Max. Explosive Likely Most Simplest and Formula Properties Gravity or Effect* Safe Safe Range Location Common Cheapest Vapor 60 Min. 8 Hour (% by vol.
  • Page 130 Isco 4700 Refrigerated Sampler Appendix B General Safety Procedures Table B-1 Hazardous Gases (Continued) Chemical Common Specific Physiological Max. Explosive Likely Most Simplest and Formula Properties Gravity or Effect* Safe Safe Range Location Common Cheapest Vapor 60 Min. 8 Hour (% by vol.
  • Page 131: Specifications

    Isco 4700 Refrigerated Sampler Index Numerics clock, internal accuracy specification, 1-5 24-hour time format, 3-3 last set report, 4-4 3-way valve control output, 3-7 setting time and date, 3-3 communications cable, 4-5 accurate volumes software, 4-5 liquid detector, 3-5 completed program, 4-2...
  • Page 132 Isco 4700 Refrigerated Sampler Index measure suction head height, 3-19 flow-paced programming, 3-12 multiple bottle compositing, 3-17 time-paced programming, 3-11 samples per bottle, 3-16 flow mode sampling, options, 3-5 suction line length, 3-4 flow pacing, 1-1 time paced sampler programming, 3-11...
  • Page 133 Isco 4700 Refrigerated Sampler Index disabled, 4-2 done, 4-2 main menu display, 3-2 halt, 4-3, 4-4 mains power cord, 1-3 pause, 4-3, 4-4 maintenance, 5-1 resume, 4-3 manual grab samples, 4-8 starting, 4-1 manual pump operation, 4-9 stop, 4-3 manual sample button, 3-2...
  • Page 134: Strainers

    Isco 4700 Refrigerated Sampler Index rinse cycles, 3-5 maximum lift, 1-4 rotor and rollers, pump, 5-3 pre-sample conditioning, 3-5 RS-232 prevent freezing, 2-14 communication cable, 4-5 routing, 2-14 serial data, 2-17 transport velocities, 1-1, 1-4 run continuously, 3-17 tubing length specification, 1-4...
  • Page 135 89/336/EEC – The EMC Directive 73/23/EEC – The Low Voltage Directive Manufacturer's Name: Teledyne Isco, Inc. Manufacturer's Address: 4700 Superior, Lincoln, Nebraska 68504 USA Mailing Address: P.O. Box 82531, Lincoln, NE 68501 Equipment Type/Environment: Laboratory Equipment for Light Industrial/Commercial Environments Trade Name/Model No:...
  • Page 136 The date of Manufacture is in code within the serial number. The first three numbers are the year of manufacture (207 is year 2007) followed by a letter for the month. "A" is January, "B" is February and so on. Hazmat Table 4700 60-4703-108 Rev.
  • Page 137 * This warranty applies to the USA and countries where Teledyne Isco Inc. does not have an authorized dealer. Customers in countries outside the USA, where Teledyne Isco has an authorized dealer, should contact their Teledyne Isco dealer for warranty service.

Table of Contents