CALIB MODE - Calibration of a RedOx probe

The RedOx probe can be re-calibrated. By nature, a RedOx probe only drifts a little and only needs to be re-calibrated in exceptional cases. This is
why RedOx calibration solutions are not supplied and their storage is not allowed to exceed 2 years. The service life of a RedOx probe is approximately
5 years. For this operation, equip yourself with the RedOx (ORP) probe, the indoor display and 240-mV and 470-mV calibration bottles. This procedure
needs to be performed in the shade (25 °C/77 °F).
1 - Put the switch in Calib mode (on the back of the indoor display).
2 - Follow the procedure
3 - Dip the end of the RedOx probe in the 240-mV calibration solution bottle for a minimum of 10 min.
A - PROCEDURE FAILED: > Check that the RedOx 240-mV solution is in contact with the probe for the duration of the procedure
> The probe no longer works (needs to be replaced).
4 - Dip the end of the RedOx probe into the 470-mV calibration solution bottle for a minimum of 10 min.
5 - The operation has been successful. > Put the switch back in ON mode
B - PROCEDURE FAILED: > Check that the RedOx 470-mV solution is in contact with the probe for the duration of the procedure
> The probe no longer works (needs to be replaced)