3410 Communication and Output settings
Figure C-4:
C.2
Ethernet cable to PC communication
The Ethernet port IP address, subnet mask, and gateway address are software-
configurable. In addition, a meter can be configured to act as a DHCP (Dynamic Host
Configuration Protocol) server to assign an IP address to a PC or laptop running Daniel
MeterLink. The DHCP server facility is not intended to act as a general purpose DHCP
server for a wider network. To this end, no user control is provided over the class or range
of IP addresses the unit provides. A standard twisted pair (Cat-5) cable should be used for
Ethernet wiring.
It is strongly recommended that the meter be configured using an independent (off-
network) single host. After configuration of the Daniel 3410 Series Gas Ultrasonic Flow
Meter, the DHCP option must be turned off if used on a LAN/WAN.
70
Optional RS-485 Module wiring
Retrofit Instructions