10.
11.
12.
13.
14.
Daniel
™
Gas Ultrasonic 3410 Series Electronics
Figure 5-8:
Base enclosure cable glands installation
Judge the proper length and cut off the excess.
Strip the outer insulation, outer shield, and inner insulation using a utility knife, refer
to Figure 5-7 for proper lengths. Verify that insulation of individual wires was not cut
while removing outer layers. Strip each wire 0.28".
Figure 5-9:
Transducer Cable Cut Instructions
Insert tubing over wires, inner insulation and under shield approximately 1.5" of
shield should overlap.
Insert and secure transducer cables through cable glands verify Line Mark is as
referenced in Figure 5-8.
Figure 5-10:
Line mark
Using a torque screwdriver set to 34 ±2 oz. in., secure the wires to the terminal plugs
of the Acquisition Module.
Installing the 3410 Series Electronics
23