Mounting and sensor wiring
April 2019
2. Gently lift the transmitter on the feed through until it disengages from the notches
3. Rotate the transmitter to the desired position.
4. Lower the transmitter, sliding it onto the notches on the feed through.
5. Replace the clamping ring on the feed through. Tighten the screw to 28 in lbf
2.6
Rotating the display orientation
The user interface orientation for the 4200 transmitter can rotate 360° in 90° increments
by software selection.
Using the display, select Menu > Configuration > Display Settings > Rotation.
18
on the feed through. You will not be able to remove the transmitter completely.
CAUTION
Do not rotate the housing more than 360°. Excessive rotation can damage the
wiring and cause measurement error or flowmeter failure.
(3.16 N m)– 32 in lbf (3.62 N m).
CAUTION
Ensure that the connection between the transmitter and the sensor is moisture-
proof. Inspect and grease all gaskets and O-rings. Moisture in the electronics can
cause measurement error or flowmeter failure.